Skip to main content
Log in/Register
  • Create new account
  • Request new password
  • Home
  • News
  • About ANKOM
  • Contact Us
ANKOM Technology logo
  • Product Catalog

    Our Product Catalog

    • Vitamin, Cholesterol & Total Fat
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI

    • Gas Fermentation Measurement
    • In Vitro Incubation
    • Filters, In Situ & Support Products
    • Chemicals And Enzymes
      • Chemicals
      • |
      • Enzymes
      • |
      • Certificates of Analysis

  • Technical Support

    Technical Support

    • FLEX Vitamin & Cholesterol
    • TDF Analyzer
    • Fiber Analyzers
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Fat Extractors
      • XT15
      • |
      • XT10

    • Hydrolysis System and RD Dryer
    • RF Gas Production System
    • Daisy Incubator
  • Analytical Procedures & Support

    Analytical Methods & Support

    • FLEX Vitamin & Cholesterol
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200

    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI
    • Gas Fermentation Measurement
    • In Vitro Incubation

  • Case Studies
    • Eurofins, Des Moines
    • Dupont Pioneer Livestock Nutrition Ctr
    • Hebron University
    • Dairy One Cooperative
    • EPL Bio Analytical Services
    • Kuwait Institute for Scientific Research
  • Global Distribution
0 items

I am getting an E11 Low Input Pressure fault during a run; what do I do?

An E11 fault is the result of the high-pressure sensor reading input pressure below 35 psi.  N2 Input pressure is always being monitored and the fault could occur at any point during a run. The fault will interrupt the run and wait for the operator to correct the problem. This problem can result from one of the following causes:

1)    The nitrogen tank has run low.
Most often this problem is because the nitrogen tank has run low and needs to be replaced, or because a valve was closed that supplied the instrument.  This could also be because the regulator was inadvertently adjusted too low and needs to be reset, up to the correct input pressure (50-55 psi gauge pressure).  Check the tank pressure and replace it if needed. Confirm that the regulator is set to 50-55 psi.  If the problem is corrected, you can resume a run in-progress.
2)    There is a substantial leak in one of the high-pressure nitrogen lines.
If there is a substantial leak in one of the high-pressure nitrogen lines, this would also cause an E11 fault. The loss of air from a leak would be audible and should be easily identified. Turn the instrument off and close the nitrogen lines. Replacement tubing, if needed, can be ordered as part 8216 Black Nitrogen Supply Tubing.  It is sold per foot and can be cut to length.
3)    The regulator is not working correctly.
It is possible that the regulator is not working correctly.  Check the regulator setting and compare it with the reading from the Pressures screen under Diagnostics.  The Diagnostics / Pressures screen displays in absolute pressure which includes atmospheric + gauge pressure.  Therefore, if the gauge pressure reads 50 psi and atmospheric pressure is 14.5 psi, the pressure displayed in the Diagnostics / Pressures screen will be 64.5 psi.  Confirm that the gauge pressure correlates with the Diagnostics/Pressures reading.  If the regulator is not adjusting pressure correctly it will not deliver the correct pressure.  If this is the case, you will need to replace part TDF61 Pressure Regulator 60 psi.
4)    The High-Pressure Sensor circuit has become shorted.
It is possible that the High-Pressure Sensor circuit has become shorted. If the black and white wires on the TDF16 High-Pressure Sensor have been shorted, a false low reading will result.  If you confirm that the correct pressure is getting to the lines and the regulator is working correctly, then the pressure sensor will need to be replaced.  Refer to Service Procedure TS002 for replacement of the TDF16 High Pressure Sensor Assembly.
 
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
 
Related Documents: 

High Pressure Regulator Replacement

AttachmentSize
PDF icon TS039_High_Pressure_Regulator_Assembly_Replacement_TDF61.pdf2.37 MB

High Pressure Sensor Replacement (TDF16)

AttachmentSize
PDF icon TS002_High_Pressure_Sensor_Replacement.pdf9.32 MB

Related Links

Contact Us
Global Distribution
Warranty Information
arrow-white-footer.pngHave Questions or Feedback?
Connect
Follow us on TwitterFind us on FacebookConnect with us on LinkedIN

Menu

  • Product Catalog
  • Technical Support
  • Analytical Methods & Support
  • Case Studies
  • Global Distribution
  • Privacy Policy
  • Refund Policy

ANKOM Technology

©2023 ANKOM Technology All Rights Reserved
ANKOM Technology 2052 O'Neil Road, Macedon NY 14502
Tel: 315.986.8090 | Fax: 315.986.8091
x

We use cookies on this site to enhance your user experience. By clicking any link on this page you are giving your consent for us to set cookies. Read our Privacy Policy.