Skip to main content
Log in/Register
  • Create new account
  • Request new password
  • Home
  • News
  • About ANKOM
  • Contact Us
ANKOM Technology logo
  • Product Catalog

    Our Product Catalog

    • Vitamin, Cholesterol & Total Fat
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI

    • Gas Fermentation Measurement
    • In Vitro Incubation
    • Filters, In Situ & Support Products
    • Chemicals And Enzymes
      • Chemicals
      • |
      • Enzymes
      • |
      • Certificates of Analysis

  • Technical Support

    Technical Support

    • FLEX Vitamin & Cholesterol
    • TDF Analyzer
    • Fiber Analyzers
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Fat Extractors
      • XT15
      • |
      • XT10

    • Hydrolysis System and RD Dryer
    • RF Gas Production System
    • Daisy Incubator
  • Analytical Procedures & Support

    Analytical Methods & Support

    • FLEX Vitamin & Cholesterol
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200

    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI
    • Gas Fermentation Measurement
    • In Vitro Incubation

  • Case Studies
    • Eurofins, Des Moines
    • Dupont Pioneer Livestock Nutrition Ctr
    • Hebron University
    • Dairy One Cooperative
    • EPL Bio Analytical Services
    • Kuwait Institute for Scientific Research
  • Global Distribution
0 items

What should I do if the enzymes are not being drawn properly?

Following are two possible reasons why enzymes may not be drawn from the reservoir.

1.      The silicone tubes may be worn or flattened from use. To inspect the tubing, refer to TDF Service Procedure TS005 Tube Replacement (Pinch Valves) and reference the steps starting at step 11, related to performing a "Valve Test."  The silicone tubes should be clear and fully opening. If any of the silicone tubes are excessively flattened or stuck in a pinched or closed position on their own (without the pistons pushing against them), then the tubes need to be replaced.  A spare set of 3 1/4" silicone tubing that was originally sent with the TDF instrument. A set of 21 replacement tubes (3 1/4") can be ordered as TDF71 Pinch Valve Tubing Set by contacting ANKOM Technology at 1-315-986-8090 or order through the Product Catalog on this website.

It is recommended that all the pinch valve tubing be replaced at the same time to keep them on the same maintenance schedule (annually). This would include the IDF, SDF, and Waste pinch valves.

2.      It is also possible that there is a blockage within the enzyme lines. Visually inspect the tubing from the enzyme reservoirs to the enzyme pinch valve to the manifold. If any crystals or dried enzymes are found, follow service procedure TS037 Enzyme Line Flush Procedure linked to below. 

NOTE: It is recommended that when running methods AOAC2009.01, 2011.25 or 2017.16 that enzyme lines be flushed after every assay using this service procedure.  If necessary, replace the enzyme supply tubing – see service procedure TS004 Tubing Replacement (enzymes).

To order an additional TDF67 Enzyme Tubing Replacement Kit contact ANKOM Technology at 1-315-986-8090 or order through the Product Catalog on this website.

Related Documents: 

Enzyme Line Flush Procedure

AttachmentSize
PDF icon TS037_Enzyme_Line_Flush_Procedure.pdf1.76 MB

Tube Replacement (Enzymes)

AttachmentSize
PDF icon TS004_Tube_Replacement_Enzymes.pdf841.65 KB

Tube Replacement (Pinch Valves)

AttachmentSize
PDF icon TS005_Tube_Replacement_Pinch_Valves.pdf2.22 MB

Related Links

Contact Us
Global Distribution
Warranty Information
arrow-white-footer.pngHave Questions or Feedback?
Connect
Follow us on TwitterFind us on FacebookConnect with us on LinkedIN

Menu

  • Product Catalog
  • Technical Support
  • Analytical Methods & Support
  • Case Studies
  • Global Distribution
  • Privacy Policy
  • Refund Policy

ANKOM Technology

©2023 ANKOM Technology All Rights Reserved
ANKOM Technology 2052 O'Neil Road, Macedon NY 14502
Tel: 315.986.8090 | Fax: 315.986.8091
x

We use cookies on this site to enhance your user experience. By clicking any link on this page you are giving your consent for us to set cookies. Read our Privacy Policy.