How do I set up my instrument for the different fiber analyses (NDF, ADF & Crude Fiber)?
How do I set up my instrument for the different fiber analyses (NDF, ADF & Crude Fiber)?
In order to properly set up your instrument for the various fiber analyses, refer to the instrument images in your A2000 Operator's Manual under the section titled Instrument Description.
Fault Screen "Cold Rinse used" "Reset . . . [ENTER]"
Fault Screen "Cold Rinse used" "Reset . . . [ENTER]"
Possible Causes:
Rinse water supply temperature did not meet method standards. This may affect results with certain samples. Make sure the water supply has a temperature no less than 50°C for Crude Fiber, or no less than 70°C for ADF and NDF determinations. This message will be displayed at the end of an analysis so your results may be affected. A small water heater dedicated to the A2000 instrument will help. Contact ANKOM for assistance as needed.