I am getting a message, “ALERT Tube (X) weak. Tubing replacement recommended;” what should I do?
I am getting a message, “ALERT Tube (X) weak. Tubing replacement recommended;” what should I do?
The ALERT message “Tube (X) weak. Tubing replacement recommended” is due to one or more of the pump tubes giving a low pressure reading during the testing of the pump tubes at the start of the analysis. When the message presents as an ALERT the problem can be ignored temporarily and addressed after the analysis is complete (Do not repeatedly ignore this message or you risk the pump tubes breaking and causing corrosion within the peristaltic pump). If the holding pressure of the pump tubes was very poor you would see an E13 Pump Tube Failure message. This ALERT is cautioning you that the pump tubes are not holding pressure well. It is possible though that if the tubing is dry (e.g if the instrument has been idle for a week) and if the lines have not been fully charged with the solutions the air in the lines can cause a weak tube message.
To diagnose the cause, first connect the TDF70 Flush Tube Assembly to all of the supply ports with the middle connector attached to the water container. Then in the Diagnostics menu, run a Line Charge (ALL) several times to ensure the lines are filled with water. Next run a Pump Tube Test, also found in Diagnostics. This will display pressure results for each of the 6 pump tubes. The minimum to PASS is a reading of at or above 8 psi. If the results are between 5.0 – 7.9 psi you will see a WARNING. If the results are less than 5.0 psi the test will display FAIL. Running this test in Diagnostics will confirm the problem, or show that is corrected with proper charging of the lines.
For more information on the weak tubes ALERT see Service Procedure TS001 E13 Fault Diagnosis. If the instrument fails the Pump Tube Test, generally the solution is to replace the pump tubes, but there can be other causes as described in TS001. Depending on the software on your instrument you would need either TDF68 Silicone Pump Tubes (software version 3.60 or earlier) or TDF99 Pump Tubes (software version 3.70 or later). Here also are procedures for installation: TS003 Service Procedure for Silicone Pump Tube Replacement and for the TDF99 Pump Tubes use TS028 Service Procedure for Long-Life Pump Tube replacement which show the location of the pump tubes and how to replace them.
I am getting an E1 Paddle Temperature Sensor Failure fault during a run; what should I do?
I am getting an E1 Paddle Temperature Sensor Failure fault during a run; what should I do?
This fault indicates that one or more of the Paddle Temperature Sensors is reporting erroneously low data of < 5oC. The fault message on the Touch Screen Display will indicate which station is having the problem and the heater to this station will be disabled. Make a note of the temperature of the paddle when this fault occurred, if possible. The remaining stations will continue to run normally. Depending on the time at which this fault occurred, one may want to disregard the test results from the suspect station.
There are three possible reasons for this fault; (1) faulty wiring from the paddle bar to the circuit board, (2) a faulty temperature sensor within the paddle bar, or (3) a faulty circuit board.
- If the E1 fault is reported for all six positions, the wiring from the paddle bar to the circuit board may be faulty. To troubleshoot, turn power off to the instrument and open the electrical enclosure. Inspect the white 3-pin connector labeled “J2” that plugs into the circuit board labeled “MUX-T”. This connector plugs in near the top of the MUX-T circuit board. Inspect the bottom of the three black wires in the J2 connector and be sure it is securely plugged into the connector. If it is loose or has become disconnected this will cause an E1 fault at all six positions. If the crimp terminal on this wire is loose or has become disconnected this will also cause this fault. If this wire appears to be securely connected into the 3-pin connector, also check to see if there is continuity from the from the same wire in the J2 connector to the other end of this wire in the 25-pin connector that the paddle bar cable plugs into. You will have to disconnect the 25-pin paddle bar connector that plugs at the bottom of the electrical enclosure to confirm this with a voltmeter. If this is identified as the problem, the TDF15 Paddle Bar Wire Harness will need to be replaced. Contact ANKOM Technology for further assistance.
- If the E1 fault is not related to all six positions then it is most likely related to the Temperature sensor within the paddle bar. When the routine in progress is done, turn the instrument off and back on again. Using the Diagnostics mode, select Temperatures and see if the station that reported the problem reading is still doing so. If the problem reading is ongoing an E1 fault will display again. If not, run a brief test. Again using the Diagnostics mode, select Heater Test, press Paddle Heater Power Enable. Having done so, turn on all six paddle heaters. Monitor the temperatures closely to see if the E1 fault returns during heating of the paddles. Do not leave the instrument at this time as the heaters will stay on potentially overheating them if not controlled. Turn the heaters off when they reach 95˚C. Allow the paddles to cool back to ambient temperature. If during the heating or cooling, the fault recurs, the Paddle Bar will need to be replaced. If the fault does not happen again, leave power on to the instrument overnight while in the Select a Function screen. The program is always monitoring temperatures when power is on and if the fault occurs again during this time, the Paddle Bar will need to be replaced. If the fault does not recur during any of these tests, it is possible that a voltage fluctuation or static discharge near the sensor may have caused this fault. The instrument may be put back into use. Keep record of any further E1 faults and if any do occur again, replace the Paddle Bar.
- It is also possible that the MUX-T circuit board is faulty. This is the least likely of the possible problems. Contact ANKOM Technology if the other options have been attempted and the problem persists.
Refer to Service Procedure TS021 Paddle Bar Replacement for replacement of the Paddle Bar Assembly (TDF23). This will require calibration of the temperature sensors and proper setting of the mixing stroke. If a new paddle or paddle bar assembly is needed, contact ANKOM Technology prior to ordering at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E2 Temperature Sensor Failure - Inline Heater fault during a run; what should I do?
I am getting an E2 Temperature Sensor Failure - Inline Heater fault during a run; what should I do?
Note: An E2 fault can be caused by cold buffer running through the instrument. If you store your buffer solutions in the refrigerator, bring them to room temperature before starting a run.
The fault message on the Touch Screen Display will indicate that the In-line Heater is disabled. The In-line Heater operates during IDF water rinses and 95% ethanol pre-heat when filling the SDF bags before the precipitation phase. The run will be automatically aborted with this fault. If a temperature of 8˚C is seen for the in-line heater, this is likely a false reading and would indicate a disconnect in the Thermistor circuit. Turn the power off and open the electrical enclosure. Identify the “MUX-T” circuit board. Locate the 3-pin connector labeled “J10” and be sure that this is securely connected to its location on the MUX-T board and that the red and white wires do not pull out of the connector easily. If the J10 connector is loose, connect it securely and see if this resolves the problem.
If the problem continues the TDF98 TDF Thermistor Assembly within the In-Line Heater will need to be replaced. If the problem persists, there may be an issue with the MUX-T board.
Note: If known, note the temperature of the In-Line Heater when this fault occurred. Depending on several factors, the SDF results from this run may not be accurate. Some samples may be more sensitive to lower temperature alcohol during the precipitation phase than others. If this fault occurred late in the 95% alcohol pre-heat / delivery, the alcohol may have nearly reached the desired temperature.
For replacement of the TDF98 TDF Thermistor Assembly see Service Procedure TS031 Thermistor Assembly Replacement to replace the thermistor.
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E3 Paddle Heater Overtemp fault; what should I do?
I am getting an E3 Paddle Heater Overtemp fault; what should I do?
To get this fault, the Temperature Sensor in one or more of the paddle heaters are reporting high temperatures. This could occur at different times under specific conditions.
- During the Amylase phase or when the instrument is idle if a temperature of over 105°C is reported.
- During the cooling cycle (after the Amylase phase) if after 20 minutes the bags have failed to drop in temperature to 62°C.
- During the AMG or Protease phase if a temperature of over 70°C is reported.
Should any one of these conditions be met during a run, the run will be aborted. The E3 error message on the Touch Screen Display will indicate which station(s) are reporting the over-temp fault. This may be an actual over-temp condition, or it may be erroneous due to a connectivity issue. Either way the heaters will be shut down by the program and the run is aborted. If this were to occur during an SDF precipitation, the heaters will be shut down, but the SDF precipitation will continue as normal.
To determine if the over-temp condition is real and the result of a heater that is failing to cycle or turn off, run the following test. Turn the power off to the instrument and allow the paddles to cool to room temperature. Now turn the power back on and from the Touch Screen Display, select Diagnostics. Press Temperatures to monitor the reported temperatures. At the same time, by touch, check the temperature of each paddle, especially the one that faulted earlier. If the suspect paddle begins getting warm, there is a problem with the paddle heater control. Turn the power off. This will require the replacement of the I/O-2 circuit board (8300). Contact ANKOM Technology for further assistance https://www.ankom.com/contact/technical-services or 1-315-986-8090.
If the previous test does not reveal a problem with the paddle(s) heating unregulated, take the following steps to identify the problem. Turn the TDF instrument power off and then back on. From the Select a Function screen, press Diagnostics and select Temperatures. See if the suspect paddle heater is again reporting an abnormally high (false) reading while sitting at room temperature. If so, there is a fault in one of the temperature sensors and the Paddle Bar Assembly needs to be replaced (TDF23) or with the newer style Paddle Bar Assembly, the individual paddles (TDF48) can be replaced. (Contact ANKOM to verify which paddle assembly you have).
If the readings appear normal and the E3 fault does not recur, run a brief test. Again, using the Diagnostics mode, select Heater Test, press Paddle Heater Power Enable. Having done so, turn on all six paddle heaters. Monitor the temperatures closely to see if the E3 fault returns during heating of the paddles. Do not leave the instrument at this time as the heaters will stay on potentially overheating them if not controlled. Turn the heaters off when they reach 90°C. Allow the paddles to cool back to ambient temperature. If during the heating or cooling, the fault recurs, the Paddle Bar will need to be replaced or with the newer style Paddle Bar Assembly the individual paddles (TDF48) can be replaced (Contact ANKOM to verify which paddle assembly you have).
If the fault does not happen again, leave power on to the instrument overnight while in the Select a Function screen. The program is always monitoring temperatures when power is on and if the fault occurs again during this time, the Paddle Bar will need to be replaced or with the newer style Paddle Bar Assembly the individual paddles (TDF48) can be replaced (Contact ANKOM to verify which paddle assembly you have). If the fault does not recur during any of these tests, it is possible that a voltage fluctuation or static discharge near the sensor may have caused this fault. The instrument may be put back into use. Keep record of any further E3 faults and if any do occur again, replace the Paddle Bar Assembly.
Refer to TDF Service Procedure #21 for replacement of the Paddle Bar Assembly (TDF23). This will require calibration of the temperature sensors and proper setting of the mixing stroke. If the problem persists or any questions arise, contact ANKOM Technology for further assistance https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E4 In-Line Heater Overtemp fault; what should I do?
I am getting an E4 In-Line Heater Overtemp fault; what should I do?
This fault occurs when the Temperature Sensor in the In-Line Heater reports a temperature exceeding 135˚C. The E4 error message on the Touch Screen Display will indicate the temperature at the time of the fault. When this fault occurs the instrument automatically aborts the current run.
The In-Line Heater operates during IDF water rinses and when filling the SDF bags with 95% alcohol. Possible causes for the E4 fault include:
1. Empty chemical container during In-line heater operation (causing an actual overheat)
This fault can occur when water or alcohol containers are not filled adequately before the run and if they empty during the In-Line Heater operation. This may cause the In-Line Heater to exceed 135˚C in which case the heater will be disabled and the run aborted. If this happens one will likely see an E14 Empty Chemical Container fault as well. Were this to happen, the instrument is performing as intended to prevent uncontrolled heating of the In-Line Heater. Be sure to start a run with containers filled to the minimum level line.
2. A short in the thermistor wiring (causing a false high reading) or a malfunctioning MUX-T circuit board (causing a false high reading)
A reading of 161˚C is reason to suspect a fault in the thermistor wiring. If this is the case, confirm that the In-line heater is not actually at this temperature. Turn the power off to the instrument and allow the In-Line Heater to cool to room temperature. Remove the clear back panel from the rear of the instrument. Turn the instrument back on and from the Select a Function screen press Diagnostics and then select Temperatures. Check to see that the In-Line Heater temperature sensor is reporting a temperature in line with the actual room temperature. If an IR thermometer is available this can be used to confirm that the In-line heater reading reflects its actual temperature. Be cautious when doing so as it could be getting very hot if there is a problem with the heater control. If the In-Line Heater is physically cool while a high temperature is reported on the display, then there is a problem with either the thermistor (TDF98 temperature sensor) or MUX-T circuit board (TDF111). Turn power off to the instrument.
3. Malfunctioning Heater Relay or I/O-2 Circuit board (causing an actual overheat)
If the In-Line Heater is actually overheating and this is confirmed both by sensing the heat coming off the heater and by a steady rise in temperature on the In-Line Heater Temperatures display – turn power off and unplug the instrument. In this case there would be a problem with either the Heater Relay (6032) or the I/O-2 Circuit board (8300). If this is the cause of the E4, inspect the in-line heater assembly for damage (i.e., severe discoloration of the brass tubes or coiled brass tubes now loose from the heater core). If damage is found, replace the entire In-Line Heater assembly.
For replacement parts and further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
To replace an In-Line Heater, click the In-Line Heater Replacement service procedure link shown here.
I am getting an E5 Paddle Heater Under Temp fault during a run; what should I do?
I am getting an E5 Paddle Heater Under Temp fault during a run; what should I do?
To get this fault, the temperature sensor in one or more of the paddle heaters is reporting low temperatures. This could occur at different times under specific conditions.
- During paddle pre-heat if any one of the paddles does not reach 50°C in 5 minutes.
- During the Amylase phase if any one of the paddle sensors does not reach 90°C in 25 minutes. Also applies to Trizma phase in the AOAC2009.01/2011.25 methods.
- During the AMG or Protease phase if any of the paddle sensors fail to maintain at least 45°C (temperature range is 60°C +10/-15°C). Also applies to Protease phase during AOAC2009.01/2011.25 methods.
Below is a list of potential causes of the E5 fault:
- Bag breakage: Confirm that an IDF or Flow-thru bag did not break during one of the heating periods. If a bag were to leak near Clamp Bar B and if fluids were to get on one of the paddles, the liquid evaporation will drop the paddle temperature and possibly cause this fault. If this has occurred, the instrument is sensing a real drop in temperature at the wetted paddle and no repair to the instrument is needed. Clean the area from the leaked bag. Test results from a broken bag should be disregarded.
- Greater volume in the IDF or Flow-thru bags: Confirm that the volume calibration is accurate. Rerun the volume calibration as described in the operator's manual (QC/Calibration Checks section). If, for example, the volume calibration is significantly off, there could be an average of 50mls of buffer delivered instead of the required 40mls. This would create a delay in reaching temperature.
- Pre-heat of paddles has been by-passed: In Diagnostics there is the option under Service Mode to "Bypass Amylase Phase Initial Heat-up." If this has been inadvertently pressed (stating "YES") this will bypass the paddle preheat to 50°C and take longer to heat up once the Buffer/Amylase is added. Make sure this option says "NO." This feature is typically used only for demo purposes.
- Clamp Bar A is left open: If for any reason Clamp Bar A was left open during the Amylase phase heat up, there will be more evaporative cooling which will require more time to heat the bags. NOTE: normally Clamp Bar A will close automatically after the final delivery to the IDF positions, but it can be re-opened.
- Front cover left open or localized draft: Be sure to use the front safety cover, especially during the Amylase phase. If this cover is left open, and/or if there is a nearby A/C vent creating a draft toward the front of the instrument the Amylase phase heat-up will take longer. Also, if the TDF instrument is located inside a hood, turn off the air flow to the hood during the Amylase phase heat-up. The continued upward airflow will create a significant delay in reaching temperature. Addressing these may resolve the problem.
- Alcohol - Buffer containers connected incorrectly: Ensure that the containers are connected correctly to the color-coded supply lines on the right side of the instrument. Confirm that alcohol did not inadvertently get put in the buffer container, and that buffer was not put in the alcohol containers. If either of these things happen, and alcohol gets dispensed to the IDF bags at the start of the Amylase phase, the alcohol will boil at a much lower temperature and not reach the target temperature of 97°C.
If the suspect paddle heater takes significantly longer to heat than the others or fails to heat up, there is a problem with either the heater within the Paddle Bar Assembly or with the I/O2 circuit board. If the Paddles heat to the touch but one of the sensors shows an erroneous low reading, there is a problem with the sensor, and the Paddle Bar will need to be replaced or with the newer style Paddle Bar Assembly the individual paddles (TDF48) can be replaced (Contact ANKOM to verify which paddle assembly you have).
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090. Follow TDF Service Procedure #TS021 for replacement instructions of the Paddle Bar assembly (TDF23).
If one of the paddles fails to heat normally, and to rule out a problem with the I/O-2 circuit board, remove the back and top panel of the Electrical Enclosure so that the I/O-2 circuit board can be seen. Do not touch wiring. Perform this simple test:
1. In Diagnostics, select Heater Test.
2. Select Paddle Heater Power Enable.
3. Check that LED #2 lights on the I/O-2 board in the electrical enclosure.
4. Leaving the above enabled, select Paddle #1, and inspect the LED (refer to chart below for each LED).
5. Repeat for each paddle heater. The following is a list of the paddles and the corresponding LED #:
- Paddle Heater Power Enable - LED 2
- Paddle 1 - LED 15
- Paddle 2 - LED 12
- Paddle 3 - LED 9
- Paddle 4 - LED 7
- Paddle 5 - LED 4
- Paddle 6 - LED 1
To test the power coming from the power supply to the individual paddle heater you will need a voltmeter set to DC. Connect the red lead on the voltmeter to the #20 position on the 24-volt DIN rail. Now turn all the heaters on and test voltage as follows. On the back of the I/O2 circuit board are two 10-pin connectors labeled CN-3 and CN-4. To test the power to each paddle, touch the black lead of the voltmeter to the following contact.
- Paddle 1 - CN4, #2 position (wire label 53)
- Paddle 2 - CN4, #5 position (wire label 50)
- Paddle 3 - CN4, #8 position (wire label 47)
- Paddle 4 - CN3, #2 position (wire label 45)
- Paddle 5 - CN3, #5 position (wire label 42)
- Paddle 6 - CN3, #8 position (wire label 39)
You should detect 25.0 volts +/- 0.2 volts at each circuit. If voltage is low there may a problem with the power supply or voltage adjustment. For details on replacing TDF110 Power Supply B see Service Procedure #TS032 Power Supply B Replacement (TDF110).
If all the LEDs are lighting up as normal and voltage is normal, the problem is with a paddle heater within the Paddle Bar Assembly. Follow Service Procedure #TS021 for replacement instructions of the Paddle Bar assembly (TDF23) or with the newer style Paddle Bar Assembly the individual paddles (TDF48) can be replaced (Contact ANKOM to verify which paddle assembly you have).
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E7 In-Line Heater Under Temp alert during a run; what should I do?
I am getting an E7 In-Line Heater Under Temp alert during a run; what should I do?
The In-line Heater operates during IDF water rinses to facilitate rinsing of the IDF bag and to ensure that solutions when deposited into the SDF bag are nominally the same temperature as the pre-heated alcohol. The run in progress will continue despite this fault. If known, note the temperature of the In-Line Heater when this fault occurred.
Note: This fault can be falsely declared if the Buffer or Water are kept in a refrigerator. The Buffer and water should be allowed to warm prior to the start of a routine.
Result: Typically, the test results from this run would not be affected by the cooler water rinse temperatures. Be sure that the sample in the IDF bag was rinsed down into the filter area of the bag and observe the precipitation phase for SDF. If any anomalies are noted pay particular attention to the test results as it is possible results for certain sample types may be more sensitive to lower temperature during the precipitation phase than others. It may be advisable to repeat the run after the problem is corrected.
Troubleshooting: After the assay with the fault is complete, turn the power off to the instrument and turn it back on to clear the fault. Test the In-line heater under Diagnostics / Heater Test. With a timer handy, observe how quickly the in-line heater heats up. Start the timer and immediately turn the In-line heater on and watch for it to get up to 70˚C. It should reach 70˚C in 40-45 seconds. Turn it off as soon as 70˚C is reached. Confirm with an IR thermometer that the temperature reading on the touch screen display matches the actual temperature of the In-line heater based on the reading taken with the IR thermometer. The two readings should be ± 5˚C of each other.
Note: If the In-line heater temperature appears to be frozen at 8˚C, there is likely a problem with a disconnected Thermistor wire at the MUX-T board or a lack of continuity in the Thermistor wiring.
If this fault is the result of a false low temperature reading, or if you need any further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
For replacement of an in-line heater, click the In-Line Heater Replacement service procedure link shown below.
I am getting an E8 In-Line Heater Under Temp alert during a run; what should I do?
I am getting an E8 In-Line Heater Under Temp alert during a run; what should I do?
Result: The In-line Heater operates during the EtOH95 delivery to the SDF bags. The run in-progress will continue despite this fault. If known, note the temperature of the In-Line Heater when this fault occurred. Typically, the test results from this run would not be affected by the cooler ethanol temperatures. Observe the precipitation phase for SDF. If any anomalies are noted pay particular attention to the test results as it is possible results for certain sample types may be more sensitive to lower temperature during the precipitation phase than others. It may be advisable to repeat the run after the problem is corrected.
Trouble: After the assay with the fault is complete, turn the power off to the instrument and turn it back on to clear the fault. Test the In-line heater under Diagnostics / Heater Test. With a timer handy, observe how quickly the in-line heater heats up. Start the timer and immediately turn the In-line heater on and watch for it to get up to 70˚C. It should reach 70˚C in 40-45 seconds. Turn it off as soon as 70˚C is reached. Confirm with an IR thermometer that the temperature reading on the touch screen display matches the actual temperature of the In-line heater based on the reading taken with the IR thermometer. The two readings should be ± 5˚C of each other.
Note: If the In-line heater temperature appears to be frozen at 8˚C, there is likely a problem with a disconnected Thermistor wire at the MUX-T board or a lack of continuity in the Thermistor wiring.
If this fault is the result of a false low temperature reading, or if you need any further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
For replacement of an in-line heater, click the In-Line Heater Replacement service procedure link shown below.
I am getting an E9 In-Line Heater Overtemp alert or fault; what should I do?
I am getting an E9 In-Line Heater Overtemp alert or fault; what should I do?
The first time this occurs during a run, the operator will have the option to “CONTINUE” or “ABORT”. If “CONTINUE” is selected the pump will resume flushing water through the in-line heater to bring the temperature down. If high temperature is reported again, the run will be aborted, and an E9 In-line Heater Overtemp Fault will be displayed.
Possible causes for the E9 fault include:
1. Empty chemical container during In-line heater cool down cycle (causing a cool-down failure)
2. A short in the thermistor wiring or a malfunctioning MUX-T circuit board
3. Malfunctioning Heater Relay or I/O-2 Circuit board
For replacement parts and further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090. For replacement of an in-line heater or MUX-T board, click the service procedure link shown below.
I am getting an E10 High Line Pressure fault during a run; what should I do?
I am getting an E10 High Line Pressure fault during a run; what should I do?
The two main reasons for this fault are:
- The pressure in the line is in fact high when the pump is delivering Buffer to IDF or EtOH95 to SDF, due to an obstruction or blockage in the fluid path.
- One of the pressure sensors could be faulty.
To analyze this problem, do the following:
- Check the pressure reading displayed when the E10 fault code appeared. If the reading was about 37 psi, this could indicate a failed pressure sensor because 37 psi is the maximum value these sensors will read and can be caused by an open circuit/disconnected wire.
- Confirm the baseline pressure readings from the six-line pressure sensors. Go to Diagnostics / Pressures and identify the readings from Lines 1 – 6. If you are at sea level the pressure should be approximately 14.2 – 15.2 psi depending on weather. The readings will vary significantly from a 14.7 psi base line due to elevation (i.e., in Denver, CO the base line pressure will be about 12.0 psi). Are the six pressure sensors reading within 0.5 psi of each other and in line with your local elevation?
If one or more of the Line pressures is significantly out of line with the others, then the TDF Pressure Sensor Assembly (TDF19) will need to be replaced.
If the six line pressures appear normal when the instrument is idle, to test further and view the dynamic pressures within each line during operation, run the following test:
E10 Fault Test:
- Obtain 6 Flow-thru bags and heat seal the bottom edges.
- Install these Flow-thru bags in the IDF positions 1-6.
- Connect the TDF70 Flush Tube Assembly to the 8 fluid ports or tubes.
- Connect the middle connector of the TDF70 Flush Tube Assembly to the Deionized Water container.
- On the Select-a-function screen select AOAC991.43 TDF.
- Start the 991.43 procedure by following the prompts on screen. There is NO need for samples, for clamp bars C&D, or for SDF bags.
- The paddles will preheat to 50°C and then fluid delivery will start to the Flow-thru bags.
- As soon as delivery begins to the Flow-thru bags press the button “View Temps/Pressures”, then press “Pressures”. Bag filling takes only about 1 minute, so you need to do this rather quickly.
- Observe the pressure readings during delivery at lines 1-6 and record these.
- When the delivery is finished you can abort the run.
What do the dynamic pressures reach during the delivery to the Flow-thru bags?
Typically, the six-line pressures will climb 1.5 – 3.5 psi above the baseline reading. So, if your baseline reading is 14.5 psi, pressure readings of 16.0 – 18.0 psi would be normal when the pump is running.
The E10 Fault is only triggered if a line pressure reaches 20.0 psi above the baseline reading. In the case of a baseline reading of 14.5 psi, the dynamic pressure would need to reach 34.5 psi.
If the pressures are climbing significantly above 18.0 psi, investigate further the source of a blockage or obstruction in the fluid path.
Here are the places that could be causing a blockage and should be inspected:
- The Pinch Valve tubing (TDF71). These should be changed annually to prevent sticking.
- The Fill Nozzles (TDF79) or Spray Tips (TDF44). Spray tips can be cleaned with the spray tip cleaning tool or replaced. The entire fill nozzle can also be replaced as needed.
- The silicone tubing within the Tubing Support Panel (TDF30). Due to the complexity of this tubing assembly, it may be necessary to replace the entire component if a blockage is suspected within this part.
- The Silicone Tubing 1/16’ x 3/16” (TDF77) that connects the Tubing Support Panel to the Fill nozzle.
- The Barbed Tube Fittings (TDF104) that attach to the vertical gusset panels near the pinch valves.
- The Brass in-line heater coils on the In-Line Heater Core (TDF120). If a blockage or corrosion cannot be flushed out of the brass tubing this component will need to be replaced.
- The Pinch Valve IDF-SDF (TDF29) may be stuck closed due to a problem with the pneumatic piston. This can be cycled and inspected after removing the tubing to confirm operation by using the Diagnostics screen / Valve Test and cycling the IDF and SDF Inlet valves.
If all the parts in the list above were to be replaced the entire flow path of the fluid from the pressure sensor assembly to the bags will have been replaced. This would address an actual blockage creating the E10 fault. If any or all these steps are taken it is advisable to rerun the E10 fault test described above.
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E11 Low Input Pressure fault during a run; what do I do?
I am getting an E11 Low Input Pressure fault during a run; what do I do?
An E11 fault is the result of the high-pressure sensor reading input pressure below 35 psi. N2 Input pressure is always being monitored and the fault could occur at any point during a run. The fault will interrupt the run and wait for the operator to correct the problem. This problem can result from one of the following causes:
I am getting an E12 High N2 Input Pressure fault during a run; what should I do?
I am getting an E12 High N2 Input Pressure fault during a run; what should I do?
2) The dial on the regulator is broken.
3) The High-Pressure Sensor has become disconnected.
4) The High-Pressure Sensor is not working.
For further assistance, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E13 Pump Tube Failure fault during a run; what do I do?
I am getting an E13 Pump Tube Failure fault during a run; what do I do?
Refer to TDF Service Procedure TS001, to diagnose a tubing pressure failure.
I am getting an E14 Empty Chemical Container fault during a run; what should I do?
I am getting an E14 Empty Chemical Container fault during a run; what should I do?
The most likely cause of an E14 fault is an empty chemical container. The fault screen will identify the container that was being drawn from when the error occurred. The vacuum sensor is designed to monitor vacuum pressure on the supply lines when they are being drawn from. When there is little or no vacuum on the line, this fault is declared. Refill the appropriate container as needed and select “Retry”.
Another potential cause of an E14 fault is a leak in any one of the supply tubes or fittings that would allow air to get pulled into a supply line. To test for this, under “Diagnostics”, run a “Motor Test" and then“Set Valves”. Open the supply line valve that generated the fault and open the output Waste valve. Set the volume to 40 mls and press “GO”. Observe the supply lines at the rear of the instrument or tilt the instrument forward using the pivot mechanism to see the supply lines. Look for any air being drawn in at a hole in the tubing or at a crack in a fitting. If this is observed, tubing or fittings may need to be replaced. For further assistance contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
Another possible cause of an E14 fault is an interruption in the Pump Motor Controller (PMC) operation. If the PMC hesitates or stops because it has overheated or because the electronics on the circuit board have been compromised, the pause in operation will cause a drop in vacuum pressure and trigger an E14 fault. This is possible even if the pause in operation were only for a second or two. You should be able to retry and continue without a problem if that is the case; however, perform further testing after the run is complete. After the run, perform a “Motor Test” under “Diagnostics” after first setting valves to Water Supply open and Waste open with all other valves closed. Set the volume to 250 mls – this will simulate the duration of operation for a 95% alcohol delivery. Fill and plug in the water container.
Press “GO” and listen for the Pump Motor operation. Be attentive to any pause in the pump’s operation. The pump should run continuously for 6.5 – 8.5 minutes, depending on the volume calibration setting for your specific instrument. Stay with the instrument for the entire delivery. Refill the water container when it nears empty, as it will draw more than 1 liter of water (250mls x 6). If no pause in operation is noticed, allow the instrument to sit idle for 10 minutes and repeat the test. If a hesitation or pause in operation is witnessed during the lengthy pump operation time the Pump Motor Controller (8301) will need to be replaced.
For further assistance contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090.
I am getting an E15 Supply Line Plugged fault during a run; what should I do?
I am getting an E15 Supply Line Plugged fault during a run; what should I do?
- Retry will simply try the delivery again.
- Override will ignore the fault for the remainder of the run.
- Abort will stop the run entirely.
To analyze the problem, do the following:
When the instrument is not in the process of running an analysis, an E15 fault can be investigated by performing a Motor Test under Diagnostics. From the Motor Test screen, press Set Valves and select the valve that caused the fault. If this is not known, test each supply valve. Open only one supply valve at a time and close all others. Open the Waste output valve and close the IDF and SDF output valves. Be sure that there is adequate fluid in the reservoir that is being tested. Change the delivery volume to 20 ml and press GO. Watch the vacuum reading on the screen. Typical readings are between 0.7 – 1.5 psi. A reading above 3 psi will generate an E15 fault (during a run). If one of the supply lines shows a high reading, you have isolated the supply line that is causing the problem.
Possible causes of an E15 fault include:
1) Clogged Filter
The most likely cause of an E15 fault at the Water, Buffer, EtOH78, EtOH95 or HCl lines is a clogged filter. Severely clogged filters may reduce delivery volumes and affect results. The filters are at the end of the draw tubes inside of the containers. These should be checked monthly and cleaned as needed. The filters can usually be cleaned with a soft bristle brush under running water. If they can't be cleaned, there are two spares of the plastic type container filters (used for water, alcohols, and HCl/Acetic Acid) and one spare of the metal screen filters (used for buffer) included with the instrument on original purchase. Additional ones may be ordered as part 8202 (plastic filter) and part 8203 (metal screen filter).
2) Quick Disconnect Fitting not fully locked in place
A quick disconnect fitting functions as a valve that closes when disconnected or not fully connected. Quick disconnect fittings are found at the following reservoirs: enzymes, water, and alcohols. Check to make sure that they cannot be pulled out without pressing the release button. If they can be pulled out without use of the release button they are not fully engaged, and fluids will not be drawn. Push them into the adjoining fitting until you hear them snap in place. Confirm that they cannot be pulled out. It is also possible that the fittings may need to be replaced – see part 8193 (coupling body) and 8194 (coupling insert) on the ANKOM website.
3) Worn tubing in the Enzyme Pinch Valve
An E15 fault at one of the Enzyme lines (Amylase, Protease, or AMG) could be due to worn tubing in the Enzyme Pinch Valve. As the tubing in the Enzyme Pinch Valve ages, it may become stuck closed preventing the flow of enzymes. Review and follow Service Procedure TS004, first replacing the tubing within the pinch valve. These tubes should be replaced annually. Spare pinch valve tubing comes with the TDF instrument upon original purchase; additional ones can be ordered as part TDF71 Pinch Valve Tubing Set (21 pcs). Also inspect the supply lines to see if there is a clog or pinch in the line. These also may require replacement. A set of all enzyme tubes can be ordered as part TDF67 Enzyme Tubing Replacement Kit. If the problem persists, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090 for further assistance.
4) Pinched or plugged supply line or a faulty supply valve
Other possible causes for an E15 fault are a pinched or plugged supply line or a faulty supply valve. Inspect the supply line in question to see if such a problem can be seen visually. After having cleaned and/or replaced the filter and having checked for fully engaged disconnect fittings, if the problem persists, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090 for further assistance.
What should be done if one of the brass supply valve is leaking or shows signs of corrosion within the valve?
What should be done if one of the brass supply valve is leaking or shows signs of corrosion within the valve?
My enzymes do not appear to be pumping (the volume in the reservoirs is staying the same); what should I do?
My enzymes do not appear to be pumping (the volume in the reservoirs is staying the same); what should I do?
The enzyme tubing may be clogged. Look at the enzyme tubing from the reservoirs to the Enzyme Pinch Valve and beyond to the Manifold. If there is any discoloration or gelled material in the tubing, replace the tubing.
Contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090 to purchase your Enzyme Tubing Replacement Kit (TDF67).
Refer to TDF Service Procedure TS004 for Enzyme Tubing installation instructions.
Why is an incorrect volume of enzyme and/or an incorrect type of enzyme being drawn after new silicone tubing is installed at the enzyme pinch valve?
Why is an incorrect volume of enzyme and/or an incorrect type of enzyme being drawn after new silicone tubing is installed at the enzyme pinch valve?
Refer to TDF Service Procedure TS004 Enzyme Tubing and reference the steps related to relaxing or pre-stressing the replacement Silicone Tubing by rolling them with a rod or dowel.
Why do the enzyme tubes appear discolored (yellow or brownish) or gelled?
Why do the enzyme tubes appear discolored (yellow or brownish) or gelled?
If enzyme tubes have become discolored (turned yellow or brownish) or if the enzyme has begun to gel in the tubes, they will require replacement. Refer to TDF Service Procedure TS004 for instructions on replacement. Contact ANKOM Technology at https://www.ankom.com/contact/technical-services or call 315-986-8090 to purchase your Enzyme Tubing Replacement Kit (TDF67).
Why am I seeing different fluid levels in the filter bags at stations 1-3 vs. 4-6?
Why am I seeing different fluid levels in the filter bags at stations 1-3 vs. 4-6?
This may be the result of one of the IDF or SDF valves being stuck in a closed or partially closed position. Using the Diagnostics screen select "Valve Test" and look for "IDF Inlet", "SDF Inlet" and "Waste" buttons. Run a brief test using the screen to open and close each valve. Confirm by looking at the pinch valve at the back of the instrument to see that it is opening and closing properly.
If any of the valves are not opening properly, refer to TDF Service Procedure TS008 to replace the pinch valve. The TDF29 Pinch Valve Assembly-IDF/SDF or TDF29.1 Pinch Valve Assembly- Waste can be ordered through the Product Catalog on this website or by calling 315-986-8090.
Little or no fluid is being delivered to the IDF and/or SDF spray nozzles at one bag position; what should I do?
Little or no fluid is being delivered to the IDF and/or SDF spray nozzles at one bag position; what should I do?
1) There could be a hole in the suspect bag position’s fluid line after or within the PUMP causing a leak in the back of the instrument. To determine if this is the cause, tilt the instrument forward and look for a leak from the silicone tubing. Open the Pump doors and check for leaks from the Pump Tubes. Close the Pump doors. Use Diagnostics to run a Pump Tube Test and check for leaks in the post-pump tubing specific to the problematic line. Observe the pressure result from the line in question. If there is a leak, a fitting or tube may need to be replaced. Contact ANKOM Technology 1-315-986-8090 for replacement parts or order replacement parts through the Product Catalog on this website. If there are no leaks but the pressure result for the suspect line is low from the previously performed Pump Tube Test, this would point to a leak prior to the Pump causing the Pump to suck air through the hole (Step 2 below) OR a blockage in the line somewhere along the suspect tube length (Step 3 below).
2) There could be a leak in the fluid line prior to the PUMP. To determine if there is a leak in the tubing line prior to the pump, run the following test:
Under Diagnostics, select Motor Test. Set Output valve Water to open and close all other Output valves and all Input valves. Set the Motor Direction to REV (reverse). Put the ends of the Waste tubes in a beaker of water but before disconnecting them from the tubing holder, label them individually 1-6 for easy identification and re-attachment later. Leave the volume set to 10 mls and press GO.
Look for any leaks prior to the pump. Follow the suspect tube back from the Pump to the Manifold looking for any leaks along the suspect line. If there is a leak, a fitting, tube, or Manifold may need to be replaced. Contact ANKOM Technology at 1-315-986-8090 for replacement parts or order replacement parts through the Product Catalog on this website. If there is no leak before or after the pump, proceed to issue step 3 below.
3) There could be a blockage in a fluid line prior to the PUMP. To identify a blockage in a line prior to the Pump, run the following test:
Under Diagnostics, select Motor Test. Set Output valve Water to open and close all other Output valves and all Input valves. Set the Motor Direction to REV (reverse). Put the ends of the Waste tubes in a beaker of water but before disconnecting them from the tubing holder, label them individually 1-6 for easy identification and re-attachment later. Next, disconnect the single (larger diameter tube) at the bottom of the Manifold from the 6-way fitting. Leave the short (1 ½”) length of tube attached to the bottom of the Manifold. Prepare to collect the fluid from the bottom of the Manifold into a clean beaker. Leave the volume set to 10 mls and press GO.
Look for any particles, dried enzyme, or mold growth being washed out of the manifold. If contaminants are seen, repeat this but draw Ethanol in at the Waste lines and repeat until the collected fluid is clear.
4) There could be a blockage in a fluid line after the PUMP. To identify a blockage in a line after the pump, consider the following. First it should be noted that by performing the procedure in step 3 above, a blockage may have been flushed backward through the line and out through the Manifold. In most cases performing the procedure in step 3 will correct a blockage before or after the pump, but if it does not, contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090 for additional diagnosing direction.
After identifying and resolving the problem, reconnect all tubing and repeat a Pump Tube Test to be sure each tube is being pressurized and holding pressure properly. Also run a Motor Test setting the Supply valve WATER to open and the IDF Output to open. Make sure the Motor Direction is FWD (forward) and place six beakers under the IDF delivery nozzles. Press GO and confirm fluid is delivered to each beaker equally. Do the same with the SDF Output valve open and beakers under the SDF delivery nozzles. If these tests pass, your instrument is ready to be put back into service.
Filtration is slow at one of the stations on my TDF instrument. How can this be corrected?
Filtration is slow at one of the stations on my TDF instrument. How can this be corrected?
Some variability in filter times is normal. Default filter time settings encompass much of this variability. Default settings for IDF/SDF filtration are as follows: IDF after release of bar B (5.0 min), IDF after water rinses (1.0 min), IDF after EtOH rinses (1.0 min), SDF after release of bar D (10.0 min), SDF after EtOH rinses 3.0 min). The following are several factors that may contribute to slow filtration (taking longer than the default settings).
Lower air pressure
Make sure that the low-pressure regulator is set properly. Our recommended setting for the low-pressure regulator is 4.0 PSI (static). Do not adjust the regulator during filtration. If your regulator is set lower than this, adjust the regulator and run the instrument again. It should also be noted that during filtration, time may be added if slow filtration is observed. As filtration is occurring, a button will appear on screen that states, "Filter Time +". Each time this is pressed, 1 minute will be added to the filtration time.
Sample type
The IDF and SDF filter bags have a significantly greater filter surface area than is found in the Gooch filter-crucible. However, regardless of filter surface area, some sample types can be difficult to filter. Samples that become quite gelatinous in an aqueous solution (such as psyllium and chia seeds) can be difficult to filter. To accommodate these types of samples in the ANKOMTDF Dietary Fiber Analyzer we recommend the following options:
Add filter time before the assay begins
At the beginning of a run, a series of user prompts appear on screen. One of these prompts states, “Filter Minutes OK? (for an IDF/SDF run) or “SDF Filter minutes OK?” for a TDF run. From this screen, you can add filter time if you are working with samples that require more filter time. Familiarity with a particular sample may dictate 10 minutes of IDF filter time, for example. Be aware that any time changes will be stored to memory and will be there for the next run. However, you will be able to change this back to the default or another setting at the beginning of each run.
Add more filter time during the filtering process
During IDF or SDF filtration and during the water or alcohol rinses (when filtering is occurring) one can press the “Filter Time +” button to add one additional minute at a time, up to 50 additional minutes. The user will need to be present at the time of filtering to take advantage of this option.
Use a smaller sample size
Reduce the sample size to 0.1 g to minimize the filter clogging effect of gelatinous sample types.
Use Diatomaceous Earth (DE) in the IDF bags
Pre-weigh 1 g of DE for each IDF bag. When starting the run, select YES for the screen that states, “Check pH manually?”. Select YES whether you plan to perform a manual pH check or not. When the instrument pauses for the pH check (and after any pH check/adjustment is complete) add 1 g of DE to each bag. Press the “Continue” button once DE is added. The DE will be mixed in during the AMG digestion. Adding the DE at this point, rather than at the beginning of the run, minimizes the risk of bag wear during mixing by having abrasive DE in a hot bag. Be sure that the DE in the IDF bags is factored into your blank values. Use of this technique helps increase filter capacity.
Wrinkles in the Filter Bags at clamp bar A or C
There is always some air leakage from the filter bags during filtration at clamp bars A and C. Larger wrinkles may exacerbate this to the point of needing to add time to the filtration. If one filter bag is taking longer than the others to filter, time may be added using the "Filter Time +" button that will be available on screen during each filtration. When setting up IDF and SDF filter bags try to keep them lying flat against the rear clamp bar before starting the run.
Clogged Spray Tip
If slow filtration is observed consistently at one individual station, inspect the spray tip at this station. The white spray tip is friction fit onto a stainless-steel tube. Remove the spray tip at the IDF or SDF station and inspect under magnification. The TDF45 Illuminated Magnifier (10x) is recommended for this. There are six small orifices (holes) around the side of the spray tip. Check to make sure they are clear and not clogged with debris. This can be cleaned with a TDF94 0.020 inch or 0.5 mm drill bit and pressurized air. You could also replace it with one of the extra spray tips that was sent with the instrument originally. After the spray tip has been cleaned or replaced turn on nitrogen to the spray tips. Using the "Diagnostics" screen, select "Valve Test" and then either "IDF N2" or "SDF N2" to turn the nitrogen on through the six spray tips. Hold a beaker of water up to each of the six spray tips to make sure nitrogen is bubbling equally at all positions. Spare spray tips can be ordered as part TDF44.
Black rubber gasket on front clamp bar A or C may be dented
Inspect the black gasket seal area at the position in question on front clamp bar A or C. See if there are any dents or low spots. The best way to check this is to lay the gasket from the front of the clamp bars against a very flat (machined) surface and see if light shows through at the position in question. If you can visually see a significant low spot or dent in the gasket at the position in question, this will contribute to a poor seal and slow filtration.
If you have a second TDF instrument, perform the following test: Switch the suspect clamp bar with the second instrument. Install filter bags at all six positions (either IDF or SDF) and close the upper and lower clamp bars keeping the bags as flat as possible to minimize leaks. Using the "Diagnostics" screen, select "Valve Test" and either "IDF N2" or "SDF N2" to inflate the bags. If the position in question now inflates its filter bag as quickly as the rest, the problem points to a dent or low spot on the original front clamp bar. Contact ANKOM Technology at https://www.ankom.com/contact/technical-services or 1-315-986-8090 for assistance. A replacement Front Clamp Bar A-position can be ordered as part TDF73; a replacement Front Clamp Bar C-position can be ordered as part TDF75.
What is the difference between IDF and SDF filter bags?
What is the difference between IDF and SDF filter bags?
SDF filter bags have a longer length of clear polypropylene material and a shorter, wider filter at the bottom of the bag.
My sample is not mixing well in one or all of my bags. How can I improve the mixing?
My sample is not mixing well in one or all of my bags. How can I improve the mixing?
If sample is mixing poorly at one or a few positions, the mixing pads may not be making good contact with the bag and paddle behind it. Refer to TDF Service Procedure TS010.
If sample is mixing poorly in general, an addition of flow control valves on the inrush and exhaust of nitrogen at the mixer piston will help. These flow control valves are accessible from the rear of the instrument and can provide better mixing control. Refer to TDF Service Procedure TS030. Also see link to part TDF113 Mixer Flow Control Valve Assembly.
What is the bag weighing procedure?
What is the bag weighing procedure?
How do I change the TDF99 Pump Tubing?
How do I change the TDF99 Pump Tubing?
Refer to TDF Service Procedure TS028 for instructions on changing the Long-Life Pump Tubing.
What are the procedures for inserting sample and Diatomaceous Earth into the filter bags?
What are the procedures for inserting sample and Diatomaceous Earth into the filter bags?
How often should I change the TDF68 Silicone Pump Tubing?
How often should I change the TDF68 Silicone Pump Tubing?
The TDF68 Silicone Pump Tubing should be changed when the instrument prompts the user with an E13 warning for weak pump tubes. The pump tubes typically last 6 months under normal use. Heavy usage may necessitate a more abbreviated timeframe between tubing changes.
Order your Pump Tubing Replacement Kit (TDF68) by contacting ANKOM Technology at 1-315-986-8090 or by ordering through the Product Catalog on this website.
Refer to TDF Service Procedure TS003 for installation instructions.
How do I perform a pH adjustment during the IDF phase?
How do I perform a pH adjustment during the IDF phase?
Follow the steps in this instructional video to perform a pH adjustment: TDF Manually Measuring pH in the IDF Digestion
Note: Measure within the appropriate pH range for your specific testing method.
What is a MoistureStop Desiccant Pouch, how is it used, and why should I use it?
What is a MoistureStop Desiccant Pouch, how is it used, and why should I use it?
The MoistureStop Desiccant Pouch is a small, airtight zipper bag, utilized for desiccating all of the Filter Bags that are used with ANKOM instrumentation. A single pouch can hold a full run of F57, F58 or XT4 bags at one time. When folded, 6 IDF or SDF bags can also be placed into a MoistureStop Desiccant Pouch. Each time a bag has been removed from the desiccant pouch, the air can and should be pushed out. The "zipper" does not have to be used with every bag removal but some action should be taken to keep the air from entering the pouch while the just removed Filter Bag is being weighed. For example, air can be pushed out of the bag by laying it on a firm surface while compressing or flattening the air out of the pouch with your hand.
Many instruments in the marketplace have the capacity to run 6 samples at a time. After being dried, the beakers, flasks or crucibles are placed into a counter top or cabinet desiccator. Once cooled, they are then removed one at a time and weighed. Each time the desiccator is opened to remove sample, moist ambient air is introduced inside the desiccator. However, because the lid or door of the desiccator is opened just 6 times, the samples that remain in the desiccator after each item is removed are affected very little by ambient moisture that may be introduced.
However, with Filter Bag Technology, generally larger numbers of Filter Bags are extracted at a time. As with the beakers, flasks and crucibles mentioned above, if the bags are placed in a desiccator after drying, each time the lid or door is opened to remove a bag, moist, ambient air is introduced. Because the desiccator is opened up to 24 times, the moisture can more readily affect the remaining bags. If a collapsible, ANKOM desiccant pouch is utilized, the air can be pushed out of the pouch each time a Filter Bag is removed. This will eliminate a build up of moisture on the remaining Filter Bags and allow for a more accurate and precise result. Every month it is possible to place the small desiccant packets into the oven at 100° - 105°C for a few hours to insure that the desiccant is renewed. In addition, to ensure the integrity of the zip lock bag, regular replacement should be considered.
How often should I change the Enzyme Tubing?
How often should I change the Enzyme Tubing?
The enzyme tubing should be changed annually under normal use. Heavy use may necessitate a shorter timeframe between tubing changes.
Order your Enzyme Tubing Replacement Kit (TDF67) by contacting ANKOM Technology at 1-315-986-8090 or by ordering through the Product Catalog on this website.
Refer to TDF Service Procedure TS004 for installation instructions.
What can cause HIGHER than expected Dietary Fiber results?
What can cause HIGHER than expected Dietary Fiber results?
Static electricity during the weighting process: The largest contributor to poor results is the effect of static electricity on the weighing process. Because plastic gloves can contribute to static electricity, only use anti-static gloves. To eliminate static electricity while the bag is being weighed, you MUST use the Bag Weigh Holder (TDF52) during the weighing process.
Moisture control: Moisture Stop desiccant pouches (X45) should be used to minimize the effect of moisture on the weighing process. Verify that the desiccant pouches are in good condition (no holes and the zipper is functioning) and the packets within the desiccant pouches are still working (they should be a blue color).
BLANK values: BLANK values are used in the IDF, SDF, and TDF calculations. If you're not using the BLANK values provided by ANKOM, try using them in your calculation.
Chemicals: Because the MES-TRIS buffer can degrade in performance due to bacteria growth, making this solution daily is important. Clean all containers daily and refill with fresh solutions. Make sure that each container is securely connected to the appropriate supply tube on the instrument.
Protein and Ash Determinations: Because protein and ash values are part of the IDF, SDF, and TDF calculations, verify that your protein and ash systems are working properly and providing expected results.
Protein and Ash preparation: See the procedures detailed in the TDF Operator's Manual (Protein and Ash Determination sections) for preparing the bags for protein and ash.
Drying: Bags must be dried at 105° C to constant weight (90 minutes).
Acetone rinses: See the procedure detailed in the TDF Operator's Manual (IDF/SDF Analysis section) for performing the acetone rinses.
Sample size: Using sample sizes larger than the recommended size of 0.5 (+/- 0.05g) can cause higher results.
Digestion times: For demonstration and diagnostic purposes, you can set the digestion times to non-standard values through the Touch Screen Display. Results will be inaccurate if the instrument is run using non-standard digestion times.
Technician variability: Confirm that all technicians are following the procedures detailed in the TDF Operator's Manual to avoid any variation in results associated with different operators.
pH: Typically the pH reading should be between 3.5 and 6 before correction. If the pH reading is outside of these values check the acid and buffer solutions, and check that tubing is connected to the proper containers.
Enzyme delivery: View the enzyme containers while the enzymes are being delivered. The fluid level should go down. If the enzymes are not being delivered properly, contact ANKOM at [email protected].
Temperature control: To verify that the temperature is being properly controlled, follow the procedure detailed in the TDF Operator's Manual (QC / Calibration section).
Mixing: To verify that the mixing is being properly controlled, follow the procedure detailed in the TDF Operator's Manual (QC / Calibration section).
Sample with bound fat: It is possible you have a sample which has been highly processed and has bound the fat to the extent the fat is not available to the solvent before the TDF process. In this case, after running the TDF instrument, remove the bags, place them on the rinse stand and perform the 2 standard acetone rinses. Immediately after the rinses (no air drying) remove the bags from the stand and soak the filters in fresh pet ether for 10 minutes. Place the bags back on the rinse stand and immediately perform 2 pet ether rinses (just like the acetone rinses). When the solvent has evaporated, seal the bags and dry them normally.
What should I do if the enzymes are not being drawn properly?
What should I do if the enzymes are not being drawn properly?
Following are two possible reasons why enzymes may not be drawn from the reservoir.
1. The silicone tubes may be worn or flattened from use. To inspect the tubing, refer to TDF Service Procedure TS005 Tube Replacement (Pinch Valves) and reference the steps starting at step 11, related to performing a "Valve Test." The silicone tubes should be clear and fully opening. If any of the silicone tubes are excessively flattened or stuck in a pinched or closed position on their own (without the pistons pushing against them), then the tubes need to be replaced. A spare set of 3 1/4" silicone tubing that was originally sent with the TDF instrument. A set of 21 replacement tubes (3 1/4") can be ordered as TDF71 Pinch Valve Tubing Set by contacting ANKOM Technology at 1-315-986-8090 or order through the Product Catalog on this website.
It is recommended that all the pinch valve tubing be replaced at the same time to keep them on the same maintenance schedule (annually). This would include the IDF, SDF, and Waste pinch valves.
2. It is also possible that there is a blockage within the enzyme lines. Visually inspect the tubing from the enzyme reservoirs to the enzyme pinch valve to the manifold. If any crystals or dried enzymes are found, follow service procedure TS037 Enzyme Line Flush Procedure linked to below.
NOTE: It is recommended that when running methods AOAC2009.01, 2011.25 or 2017.16 that enzyme lines be flushed after every assay using this service procedure. If necessary, replace the enzyme supply tubing – see service procedure TS004 Tubing Replacement (enzymes).
To order an additional TDF67 Enzyme Tubing Replacement Kit contact ANKOM Technology at 1-315-986-8090 or order through the Product Catalog on this website.
What can cause LOWER than expected Dietary Fiber results?
What can cause LOWER than expected Dietary Fiber results?
Static electricity during the weighting process: The largest contributor to poor results is the effect of static electricity on the weighing process. Because plastic gloves can contribute to static electricity, only use anti-static gloves. To eliminate static electricity while the bag is being weighed, you MUST use the Bag Weigh Holder (TDF52) during the weighing process.
Sample sticking to the IDF Flow-thru bags: When running a TDF assay, particles can stick to the IDF Flow-thru bag and not get rinsed down into the SDF filter. The weight of the sample stuck to the IDF Flow-thru bag is not accounted for in the TDF calculation. To avoid this problem, after the digestion processes are complete, use 78% EtOH to manually rinse any sticky particles down from the IDF Flow-thru bag into the SDF bag. If necessary, a laboratory spatula can be used to help loosen particles that are stuck to the bag.
BLANK values: BLANK values are used in the IDF, SDF, and TDF calculations. If you're not using the BLANK values provided by ANKOM, try using them in your calculation.
Chemicals: Because the MES-TRIS buffer can degrade in performance due to bacteria growth, making this solution daily is important. Clean all containers daily and refill with fresh solutions. Make sure that each container is securely connected to the appropriate supply tube on the instrument.
Protein and Ash Determinations: Because protein and ash values are part of the IDF, SDF, and TDF calculations, verify that your protein and ash systems are working properly and providing expected results.
Protein and Ash preparation: See the procedures detailed in the TDF Operator's Manual (Protein and Ash Determination sections) for preparing the bags for protein and ash.
Drying: Bags must be dried at 105°C to constant weight (90 minutes).
Acetone rinses: See the procedure detailed in the (IDF/SDF Analysis section) for performing the acetone rinses. TDF Operator's Manual
Sample loss: If any sample is spilled or left in a weigh tin during transfer to the IDF bag, the assay will produce lower values.
Filtrate transfer from IDF to SDF bag: If any liquid does not quantitatively transfer from the IDF bag to the SDF bag, SDF results will be low. Make sure that the bottom of the IDF bags remain inside the top of the SDF bags during the IDF filter process.
Diatomaceous Earth (DE) loss: If any DE is spilled or left in a weigh tin during transfer to the SDF bag, the assay will produce lower values.
Digestion times: For demonstration and diagnostic purposes, you can set the digestion times to non-standard values through the Touch Screen Display. Results will be inaccurate if the instrument is run using non-standard digestion times.
Technician variability: Confirm that all technicians are following the procedures detailed in the TDF Operator's Manual to avoid any variation in results associated with different operators.
pH: Typically the pH reading should be between 3.5 and 6 before correction. If the pH reading is outside of these values check the acid and buffer solutions, and check that tubing is connected to the proper containers.
How do I replace a broken Fill Nozzle?
How do I replace a broken Fill Nozzle?
Refer to TDF Service Procedure TS014 Fill Nozzle Installation Procedure.
How do I prepare bags for Kjeldahl & Ashing procedures?
How do I prepare bags for Kjeldahl & Ashing procedures?
In order to properly prepare the IDF & SDF bags for Kjeldahl and Ashing procedures, view the documents below or watch the following video: Kjeldahl & Ashing Bag Preparation.
Nitrogen amount and purity?
Nitrogen amount and purity?
Do NOT run this instrument without an adequate DRY Nitrogen Supply. The instrument is powered by pneumatics that are not compatible with moisture, commonly found in compressed air. Nitrogen is also recommended as an extra precaution when handling ethanol. For these reasons Compressed Air is NOT recommended.
Amount of Nitrogen
A typical TDF run requires approx. 1400L (50cu.ft.) nitrogen gas. The nitrogen supply should be 6.9 bar (100 psi). During the run, the nitrogen is used to operate pressure cylinders and also to facilitate bag filtering. During the filtering stage, the nitrogen is regulated to .3 bar (4 psi) and flow rate is approx. 15L/min (.5cu.ft.).
Nitrogen Purity
A key issue associated with the gas used in the TDF instrument is moisture content. The TDF instrument can suffer damage if moisture is introduced through the gas line. ANKOM recommends the use of nitrogen with a 99.998% purity rating. Theoretically, less pure nitrogen should work if the impurities do NOT include moisture.
What times and temperatures should I set on my Kjeldahl to determine the protein in the IDF/SDF filter bags?
What times and temperatures should I set on my Kjeldahl to determine the protein in the IDF/SDF filter bags?
How much bench space does the ANKOM TDF Dietary Fiber Analyzer require and how much does it weigh?
How much bench space does the ANKOM TDF Dietary Fiber Analyzer require and how much does it weigh?
For shipment, some parts are removed and the dimensions are: 107 cm long (42") x 51 cm deep (20") x 76 cm high (30").
The TDF instrument alone weighs about 65 Kgs (142 lbs). With the controller, attachments, and other spare parts the weight is about 71 kgs (157 lbs). A rinse stand adds 2.75 kgs (6 lbs) to the shipping weight. The total weight for shipping purposes, when the order includes a rinse stand, is 93.5 kgs (206 lbs).
Is there a recommended procedure for storing the TDF Instrument during a period of non-use?
Is there a recommended procedure for storing the TDF Instrument during a period of non-use?
Yes, there is a recommended procedure for the TDF Instrument for periods of non-use in Appendix B of the Operator's Manual for instructions. Following this procedure will help maintain clear fluid lines.
The most current Operators Manual can be found at https://www.ankom.com/technical-support/tdf-analyzer?f%5B0%5D=field_question_subtype%3A20
Are there any special instructions for the return of a TDF Analyzer for service?
Are there any special instructions for the return of a TDF Analyzer for service?
Yes. Please refer to TDF Service Procedure TS019 (link below) for details on how to ship your TDF Analyzer to ANKOM Technology.
What is the recommended maintenance for the TDF instrument?
What is the recommended maintenance for the TDF instrument?
You can access these documents using the links below.
How do I handle samples that are difficult to filter?
How do I handle samples that are difficult to filter?
Add filter time before the assay begins
At the beginning of a run a series of user prompts appear on screen, one of which states, “Filter Minutes OK? (for an IDF/SDF run) or “SDF Filter minutes OK?” for a TDF run. From this screen, you can add filter time if you are working with samples that require more filter time. Familiarity with a particular sample may dictate 10 minutes of IDF filter time, for example. Be aware that any time changes will be stored to memory and will be there for the next run. You will, however, be able to change this back to the default or another setting at the beginning of each run.
Add more filter time during the filtering process
During IDF or SDF filtration and during the water or alcohol rinses (when filtering is occurring) one can press the “Filter Time +” button to add one additional minute at a time, up to 50 additional minutes. The user will need to be present at the time of filtering to take advantage of this option.
Use a smaller sample size
Reduce the sample size to 0.1 g to minimize the filter clogging effect of gelatinous sample types. NOTE: Remember, the TDF values will be greatly influenced by variability in protein and ash values - of a small sample size.
Use Diatomaceous Earth (DE) in the IDF bags
Use of this technique helps increase filtering capacity. Pre-weigh 0.5-1 g of DE for each IDF bag. About 2 min before the end of AMG digestion, open Clamp Bar A by pressing "Clamp Bar A", add the DE quantitatively, and then close Clamp Bar A by pressing "Clamp Bar A" again. The DE will be mixed in during the end of the AMG digestion. Adding DE at this point, rather than at the beginning of the run, minimizes the risk of bag wear from abrasive DE in a hot/agitated bag. Be sure that the DE in the IDF bags is factored into your blank values. We recommend running a blank position when implementing this technique to account for DE loss from the IDF bag, if there are any.
What are the filtration specifications for the ANKOM TDF filtration process?
What are the filtration specifications for the ANKOM TDF filtration process?
How do I analyze liquid samples in the TDF Analyzer?
How do I analyze liquid samples in the TDF Analyzer?
- Weigh the pipette and tare the weight.
- Fill the pipette with a liquid sample.
- Place the pipette on the balance and record the weight.
- Pipette out the liquid into the IDF Bag.
- Weigh the pipette and subtract that weight from the weight with a sample. This supplies us with the actual sample weight that went into the bag.
I am getting high TDF results even after doing the de-fat procedure. How can I correct this?
I am getting high TDF results even after doing the de-fat procedure. How can I correct this?
- During the de-fat procedure, do two 10 minute soaks in petroleum ether. This is particularly helpful with high fat (> than 20%) samples.
- It is possible you have a sample which has been highly processed and has bound the fat to the extent the fat is not available to the solvent before the TDF process. In this case, after running the TDF instrument, remove the bags, place them on the rinse stand and perform the 2 standard acetone rinses. Immediately after the rinses (no air drying) remove the bags from the stand and soak the filters in fresh pet ether for 10 minutes. Place the bags back on the rinse stand and immediately perform 2 pet ether rinses (just like the acetone rinses). When the solvent has evaporated, seal the bags and dry them normally.
I see a white substance floating on top of the sulfuric acid after a Kjeldahl digestion. Will this substance affect my results?
I see a white substance floating on top of the sulfuric acid after a Kjeldahl digestion. Will this substance affect my results?
How frequently should the silicone delivery tubing be replaced?
What waste is produced by the TDF instrument?
What waste is produced by the TDF instrument?
My enzymes are draining out of the reservoirs when the instrument is idle, what should I do?
My enzymes are draining out of the reservoirs when the instrument is idle, what should I do?
For instruments with version 3.78 or newer software perform the following diagnostics routine:
1.
|
Run Line Charge routine two times with “All” selected with the flush tube assembly installed using deionized water. This will prime the lines in the instrument.
|
2.
|
Set up Motor test with the Flush Tubes still connected.
|
a. Set all Supply Valves to Closed
b. Set Speed to 240 RPM
c. Leave Amount set to 10 ml
d. Press Start
|
|
3.
|
At the start of the routine, in the upper right corner of the screen you should see a number approximately 14. Part way through the routine you should see it say “Plugged”. This is normal for this stage of the test.
|
4.
|
Below the word “Plugged” you will see a set of numbers that will climb throughout the routine and should end in the range of 11 or higher. Any number below 11 indicate a leaky valve. In most cases, the valve causing the issue will be the buffer valve.
|
5.
|
Refer to Service Procedure TS-017 for maintenance instructions for the valve. If the valve body is pitted, especially around the hole in the center, it will be necessary to replace the valve body using the TDF25.5 kit. Follow Service Procedure TS046 for installation instructions.
|