Why is an incorrect volume of enzyme and/or an incorrect type of enzyme being drawn after new silicone tubing is installed at the enzyme pinch valve?

Refer to TDF Service Procedure Tube Replacement Enzymes (TS004) and review the steps related to relaxing or pre-stressing the replacement silicone tubing by rolling them with a rod or dowel.