The Chamber is full of liquid at the end of the Hydrolysis Procedure. What should I do?

Possible Causes:

(1) The vent tube or drain tube is blocked. The blockage is caused by water trapped in the line. It is very important that the vent and drain tubes run downhill and are not submerged in water. If the vent tube is blocked, a vacuum is created, and the vessel cannot drain.

(2) There is a faulty drain valve. If tubes are functioning properly, contact ANKOM for instruction in testing the drain valve.