Fault Screen "Vessel Overtemp"

The vessel temperature has exceeded the maximum limit.

Possible Causes:

The solution level is low in the vessel, but the level sensor is indicating full. The level sensor has two main components that work together. These components are a Level Sensor Module and the Glass Sensor Tip. Refer to and follow the section on testing and training the Level Sensor in A2000 Service Procedure 10 for SICK Level Sensor (Blue Model) or see A2000 Service Procedure 7 for Analog Level Sensor (Black Model).

If the level sensor is not working, it may need replacing. See A2000 Service Procedure 11 for the SICK Level Sensor (Blue Model) or see A2000 Service Procedure 9 for the Analog Level Sensor Module (Black or gray model) to replace the Glass Tip and Fiber Optic Cable Assembly (A32).

The vessel is overfilled with solution. The Level Sensor is not detecting solution. See A2000 Service Procedure 10 for SICK Level Sensor (Blue Model) or see A2000 Service Procedure 7 for Analog Level Sensor (Black Model).

The motor stopped turning. Check the Packing Nut to make sure it is not over tightened. See A2000 Service Procedure 1.

The motor needs to be replaced. See A2000 Service Procedure 23.