Skip to main content

FAQ

What can I do to extend battery life?

Battery life can be extended by unplugging the battery (Part #: RF16 or RF46) when a module is not being used. In addition, utilizing a longer live interval, such as 60 seconds, will extend battery life. Minimizing the use of the "Open Valves" button or "Valve Open" check boxes will also conserve battery life. This feature has the most significant drain on the battery as this holds the valve open requiring a constant draw on the battery. 

Unless it is important to have the valve held open, such as for cleaning the vent valve tubing, use the "Global Release" set to "0" to bring the pressure to zero at the beginning of a test.