Possible Causes: Possible Causes: Leakage has occurred at the agitator shaft; therefore, the packing nut requires attention. See DELTA Service Procedure Agitator Packing Maintenance and Nut Adjustment (DES001). If tightening the packing nut does not...
Possible Causes: The packing nut may be too tight. For proper tightening instructions, refer to DELTA Service Procedure Agitator Packing Maintenance and Nut Adjustment (DES001). The packing assembly may have corrosive buildup. If periodic maintenance...
Possible Causes: (1) The bag suspender is worn or not properly functioning. Refer to DELTA Service Procedure Bag Suspender Check (DES006). (2) The level sensor is not detecting the solution. Refer to DELTA Service Procedure Level Sensor (Analog)...
See DELTA Service Procedure Agitation System Check (DES031).