In the event of a Drain Time-Out Fault (MF7) or an Abort during a Drain Module, users must flush the lines beneath the vessels to clear the residue prior to starting a new run. For more information, refer to Section 17 of the FLEX Operator’s Manual...
This fault indicates a System Integrity Failure. One or more columns or round bottom flasks did not reach or hold pressure properly. To resolve an MF24 fault: 1. Confirm the Low N2 setting is: 2.6 - 3.0 PSI. See Section 6.7.2 of the FLEX Operator's...
This fault indicates a System Integrity Failure. One or more vessels did not reach or hold pressure properly. To resolve an MF1 fault: 1. Confirm the Low N2 setting is: 2.6 - 3.0 PSI. See Section 6.7.2 of the FLEX Operator’s Manual. 2. Tighten the...