An E11 fault is the result of the high-pressure sensor reading input pressure below 35 psi. N2 input pressure is always monitored and the fault could occur at any point during a run. The fault will interrupt the run and wait for the operator to...
Refer to TDF Service Procedure E13 Diagnosis (TS001) to diagnose tubing pressure failure.
The touchscreen display will give you the option to Retry , Override , or Abort the run. Retry will simply try the delivery again. Override will ignore the fault for the remainder of the run. Abort will stop the run entirely. To analyze the problem...
This fault indicates that one or more of the Paddle Temperature Sensors is reporting erroneously low data of < 5°C. The fault message on the Touch Screen Display will indicate which station is having the problem and the heater to this station will be...
The touchscreen display will give you the option to Retry , Override , or Abort the run. Retry will simply try the delivery again. Override will ignore the fault for the remainder of the run. Abor t will stop the run entirely. Possible causes include...
Cause: This fault occurs when the temperature sensor to the in-line heater reports that the in-line heater is not reaching the required 80˚C within the 90 seconds allotted during an EtOH95 delivery. The fault message on the touchscreen display will...
An E10 fault is the result of one or more of the six pressure sensors reading a pressure 20 psi above the baseline pressure during the buffer delivery to the IDF bags or during the EtOH 95 delivery to the SDF bags. This fault will abort the run. The...
To get this fault, the temperature sensor in one or more of the paddle heaters is reporting high temperatures. This could occur at different times under specific conditions such as: During the Amylase phase or when the instrument is idle if a...
This fault occurs when the temperature sensor in the in-line heater reports a temperature exceeding 135˚C. The E4 error message on the touch screen display will indicate the temperature at the time of the fault. When this fault occurs the instrument...
This fault occurs when the temperature sensor reports a high temperature while the in-line heater is attempting to cool within the required timeframe. Cooling is facilitated by flushing water through the lines of the in-line heater. The first time...
An E12 fault is the result of the high-pressure sensors reading input pressure above 100 psi. N2 Input pressure is always monitored, and the fault could occur at any point during a run. The fault will interrupt the run and wait for the operator to...
To get this fault, the temperature sensor in one or more paddle heaters is reporting low temperatures. This could occur at different times under specific conditions such as: During paddle pre-heat, if any one of the paddles does not reach 50°C in...
Cause: This fault occurs when the temperature sensor in the in-line heater reports that the in-line heater is not reaching the required 110˚C within the 90 seconds allotted during an IDF water rinse. The fault message on the touchscreen display will...
An E2 Fault is declared when the thermistor to the in-line heater is reporting an erroneously low reading of < 10˚C. IMPORTANT: An E2 fault can be caused by cold buffer running through the instrument. If you store your buffer solutions in the...