When the instrument is turned on, the drain valve opens to vent the vessel. If the drain open/close switch is not sensed by the computer, this fault occurs. Possible Causes: The wires on the drain valve have dislodged. Connections may have become...
Possible Causes: (1) The level sensor has failed to detect the liquid that is in the vessel. If the vessel is full and this fault screen appears, press <STOP> to drain the fluid. You may want to capture the solution and re-use it. Refer to DELTA...
A drain error usually occurs if the level sensor is sensing solution in the vessel after the drain has opened. If the vessel is empty, and the level sensor indicates "Full," then the level sensor must be fixed. If the vessel is full of solution after...
Possible Causes: If liquid has been manually poured into the vessel before the computer communicates with the appropriate valve to fill the vessel, this fault screen will appear. Open the vessel, if the vessel is full press "Y<ENTER>" to continue the...