The XT instrument’s flowmeter should have a flow rate between 4-6 gallons per hour (If chilled coolant is used then the flow rate should be 2-3 gallons per hour). Note that this flow is intermittent and only turns on when cooling is required. When...
When using the Mojonnier mixture of solvents, a black residue may build up on the bottom of the vessel. If this coating becomes too thick, an "Overtemp Fault" "<Enter>" may occur. The black residue can be cleaned using a drill with a long-shafted...
Refer to XT15 Service Procedure Changing the Solvent (X5S007).
The site glass assembly can be removed and cleaned; however, it is highly recommended to replace the tubing to the site glass after cleaning. Contact ANKOM Technical Service at https://www.ankom.com/contact/technical-services for instructions. The...
Refer to XT15 Service Procedure Instrument and Components Cleaning (X5S017).
Below is a list of steps to verify that the XT instrument is operating correctly. Always refer to your manual for complete, detailed information. Refer to XT15 Service Procedure Solvent, Vent, and Water Diagram (X5S006) for a four-page diagram that...