Skip to main content
Log in/Register
  • Create new account
  • Request new password
  • Home
  • News
  • About ANKOM
  • Contact Us
ANKOM Technology logo
  • Product Catalog

    Our Product Catalog

    • Vitamin, Cholesterol & Total Fat
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI

    • Gas Fermentation Measurement
    • In Vitro Incubation
    • Filters, In Situ & Support Products
    • Chemicals And Enzymes
      • Chemicals
      • |
      • Enzymes
      • |
      • Certificates of Analysis

  • Technical Support

    Technical Support

    • FLEX Vitamin & Cholesterol
    • TDF Analyzer
    • Fiber Analyzers
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Fat Extractors
      • XT15
      • |
      • XT10

    • Hydrolysis System and RD Dryer
    • RF Gas Production System
    • Daisy Incubator
  • Analytical Procedures & Support

    Analytical Methods & Support

    • FLEX Vitamin & Cholesterol
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200

    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI
    • Gas Fermentation Measurement
    • In Vitro Incubation

  • Case Studies
    • Eurofins, Des Moines
    • Dupont Pioneer Livestock Nutrition Ctr
    • Hebron University
    • Dairy One Cooperative
    • EPL Bio Analytical Services
    • Kuwait Institute for Scientific Research
  • Global Distribution
0 items

Fault Screen "Vessel Overtemp"

The vessel temperature has exceeded the maximum limit.

Possible Causes:

The solution level is low in the vessel, but the level sensor is indicating full. The level sensor has two main components that work together. These components are a Level Sensor Module and the Glass Sensor Tip. Refer to and follow the DELTA Service Procedure 007 for Analog Level Sensor Cleaning and Testing.

If the level sensor is not working, it may need replacing. See DELTA Service Procedure 009 to replace the Analog Level Sensor Glass Tip and Fiber Optic Cable (A32).

The vessel is overfilled with solution. The Level Sensor is not detecting solution. See DELTA Service Procedure 007 for Analog Level Sensor Cleaning and Testing.

The motor stopped turning. Check the Packing Nut to make sure it is not over tightened. See DELTA Service Procedure 001.

The motor needs to be replaced. See DELTA Service Procedure 023.

Related Documents: 

Agitator Packing Maintenance and Nut Adjustment

AttachmentSize
PDF icon DES001_Agitator_Packing_Maintenance_and_Nut_Adjustment.pdf698.31 KB

Level Sensor Analog Cleaning and Testing

AttachmentSize
PDF icon DES007_Level_Sensor_Analog_Cleaning_and_Testing.pdf716.8 KB

Level Sensor Analog Cable and Glass Tip Replacement

AttachmentSize
PDF icon DES009_Level_Sensor_Analog_Cable_and_Glass_Tip_Replacement.pdf269.94 KB

Motor Assembly Replacement

AttachmentSize
PDF icon DES023_Motor_Assembly_Replacement.pdf599.91 KB

Related Links

Contact Us
Global Distribution
Warranty Information
arrow-white-footer.pngHave Questions or Feedback?
Connect
Follow us on TwitterFind us on FacebookConnect with us on LinkedIN

Menu

  • Product Catalog
  • Technical Support
  • Analytical Methods & Support
  • Case Studies
  • Global Distribution
  • Privacy Policy
  • Refund Policy

ANKOM Technology

©2023 ANKOM Technology All Rights Reserved
ANKOM Technology 2052 O'Neil Road, Macedon NY 14502
Tel: 315.986.8090 | Fax: 315.986.8091
x

We use cookies on this site to enhance your user experience. By clicking any link on this page you are giving your consent for us to set cookies. Read our Privacy Policy.