Skip to main content
Log in/Register
  • Create new account
  • Request new password
  • Home
  • News
  • About ANKOM
  • Contact Us
ANKOM Technology logo
  • Product Catalog

    Our Product Catalog

    • Vitamin, Cholesterol & Total Fat
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI

    • Gas Fermentation Measurement
    • In Vitro Incubation
    • Filters, In Situ & Support Products
    • Chemicals And Enzymes
      • Chemicals
      • |
      • Enzymes
      • |
      • Certificates of Analysis

  • Technical Support

    Technical Support

    • FLEX Vitamin & Cholesterol
    • TDF Analyzer
    • Fiber Analyzers
      • DELTA
      • |
      • A2000
      • |
      • A200
    • Fat Extractors
      • XT15
      • |
      • XT10

    • Hydrolysis System and RD Dryer
    • RF Gas Production System
    • Daisy Incubator
  • Analytical Procedures & Support

    Analytical Methods & Support

    • FLEX Vitamin & Cholesterol
    • Dietary Fiber
    • Crude/Detergent Fiber
      • DELTA
      • |
      • A2000
      • |
      • A200

    • Total/Crude Fat
      • XT15
      • |
      • XT10
      • |
      • HCI
    • Gas Fermentation Measurement
    • In Vitro Incubation

  • Case Studies
    • Eurofins, Des Moines
    • Dupont Pioneer Livestock Nutrition Ctr
    • Hebron University
    • Dairy One Cooperative
    • EPL Bio Analytical Services
    • Kuwait Institute for Scientific Research
  • Global Distribution
0 items

How do I handle samples that are difficult to filter?

Some sample types can be difficult to filter, even though the IDF and SDF filter bags have a significantly greater filter surface area than the Gooch crucible. Some samples become gelatinous in an aqueous solution (such as psyllium, chia seeds, or milled beans) and can be difficult to filter.  To accommodate these types of samples, we recommend the following options:

Add filter time before the assay begins
At the beginning of a run a series of user prompts appear on screen, one of which states, “Filter Minutes OK? (for an IDF/SDF run) or “SDF Filter minutes OK?” for a TDF run.  From this screen, you can add filter time if you are working with samples that require more filter time.  Familiarity with a particular sample may dictate 10 minutes of IDF filter time, for example. Be aware that any time changes will be stored to memory and will be there for the next run.  You will, however, be able to change this back to the default or another setting at the beginning of each run.

Add more filter time during the filtering process
During IDF or SDF filtration and during the water or alcohol rinses (when filtering is occurring) one can press the “Filter Time +” button to add one additional minute at a time, up to 50 additional minutes.  The user will need to be present at the time of filtering to take advantage of this option.

Use a smaller sample size
Reduce the sample size to 0.1 g to minimize the filter clogging effect of gelatinous sample types. NOTE: Remember, the TDF values will be greatly influenced by variability in protein and ash values - of a small sample size.

Use Diatomaceous Earth (DE) in the IDF bags 
Use of this technique helps increase filtering capacity. Pre-weigh 0.5-1 g of DE for each IDF bag.  About 2 min before the end of AMG digestion, open Clamp Bar A by pressing "Clamp Bar A", add the DE quantitatively, and then close Clamp Bar A by pressing "Clamp Bar A" again. The DE will be mixed in during the end of the AMG digestion. Adding DE at this point, rather than at the beginning of the run, minimizes the risk of bag wear from abrasive DE in a hot/agitated bag.  Be sure that the DE in the IDF bags is factored into your blank values. We recommend running a blank position when implementing this technique to account for DE loss from the IDF bag, if there are any.

Related Links

Contact Us
Global Distribution
Warranty Information
arrow-white-footer.pngHave Questions or Feedback?
Connect
Follow us on TwitterFind us on FacebookConnect with us on LinkedIN

Menu

  • Product Catalog
  • Technical Support
  • Analytical Methods & Support
  • Case Studies
  • Global Distribution
  • Privacy Policy
  • Refund Policy

ANKOM Technology

©2023 ANKOM Technology All Rights Reserved
ANKOM Technology 2052 O'Neil Road, Macedon NY 14502
Tel: 315.986.8090 | Fax: 315.986.8091
x

We use cookies on this site to enhance your user experience. By clicking any link on this page you are giving your consent for us to set cookies. Read our Privacy Policy.