Some samples with large amounts of carbohydrates (starches and cellulose) may bind small quantities of solvent and require additional post drying. To remove the bound solvent, dry at 100-105° C for one to three hours.
To perform an extraction of oilseeds, refer to Oilseed Extractions (FOSFA Equivalent).
Research with high fat samples has given us the following understandings: Some samples with fats that have low melting point triacylglycerides or emulsifiers may encounter fat migration during hydrolysis. To improve fat retention, increase the level...
High Fat Samples are defined as any meat sample with a fat content >20% and any plant sample with a fat content >40%. See XT Service Procedure High Fat Sample Handling (XTS014).
Replacement of the dryer completely depends upon how often it is used to dry hydrolysis samples. The filter should be changed when the filter bags do not dry sufficiently after hydrolysis at 110°C for three hours (adding pH paper in the desiccant...
When using the ANKOM Rapid Dryer to dry acid hydrolysis samples, the airflow through the special hydrolysis filter ( Part #: RD2 ) will eventually become too limited and the filter will need to be changed. To check the air filter for proper airflow...
When samples are dried for three hours after hydrolysis, they should be placed in a desiccant pouch ( Part #: X45 ) with a strip of pH paper. When the bags are fully dry, the pH paper will NOT show the presence of acid.
This can occur with samples that are very low in fat. Moisture left in low-fat samples after extraction is the issue. As AOCS Procedure AM 5-04 indicates, when the Filter Bags are removed from the Extractor, they should be placed in an oven at 100...
You can see unexpectedly high values when hydrolyzing fat samples if the samples are not adequately dried after the hydrolysis step. Possible reasons for insufficient drying include: (1) Dryer is not working properly - Filter needs to be replaced -...
Possible Causes: (1) The vent tube or drain tube is blocked. The blockage is caused by water trapped in the line. It is very important that the vent and drain tubes run downhill and are not submerged in water. If the vent tube is blocked, a vacuum is...
Possible Causes: (1) The dial on the heat sealer may not be turned up high enough. To seal the XT4 Bag, the dial must be set at approximately "6." (2) The heating element is broken. On the sides of the Teflon cover are two shiny metal strips that...
Refer to the Fat Extraction Process section of the ANKOM XT10 Extractor Operator's Manual. See subsection, Sample Preparation Procedure for Crude Fat.
Yes. For instructions on how to extract liquid samples, view ANKOM XT10 & XT15 Crude (Free) Fat Extraction of Liquid Samples and refer to the Sample Preparation section of the Operator’s Manual.
Yes, all samples should be treated the same during the drying process. The biggest and most common cause of artificially high crude fat results is incomplete drying in the pre-extraction phase. Because our technique uses the indirect method of...
When the lid is raised, some residual acid may flow onto the hinges and the handle. Over time, corrosion will occur if this happens. When lifting the lid after a run, it is recommended that the lid be wiped off with a paper towel to prevent dripping...
If a Mojonnier solvent (pet ether, ethyl ether, ethanol) is utilized, a blank value of 0.0030g is normal. If pet ether or hexanes are used, a blank value of 0.0010g is normal. During the hydrolysis phase, fat migration may occur with samples that...
The XT instrument’s flowmeter should have a flow rate between 4-6 gallons per hour (If chilled coolant is used then the flow rate should be 2-3 gallons per hour). Note that this flow is intermittent and only turns on when cooling is required. When...
Crude Fat results could be low for any of the following reasons: The extraction was not run long enough. The sample was not post-extraction dried enough. Fat migrated from a bag during the pre-dry step. Desiccant was either not used or incorrectly...
To determine how much DE to use with various sample types, refer to the Hydrolysis Procedure subsection titled Sample Categorization of the HCl Operator's Manual.
The MoistureStop Desiccant Pouch is a small, airtight zipper bag, utilized for desiccating all of the Filter Bags that are used with ANKOM instrumentation. A single pouch can hold a full run of F57, F58 or XT4 bags at one time. When folded, 6 IDF or...
Petroleum ether is the most common solvent recommended for use in XT Extractors. Many other solvents will work depending on your objectives but may require upgrades to the method or instrument for them to run effectively. Hexane, for example, will...
This is likely due to a valve that may be "stuck" closed. If the instrument has not been used in a while, it is possible that the solution is simply to clean the valves and clean any debris or buildup that may have settled on the valve piston. To...
There is a leak. The most common cause is a leaky Q1 valve, which would cause cooling water to flow into the vessel during the heating stage. To test: Turn the instrument completely off but keep the water supply on. Check to see if water is running...
The XT10 and XT15 look very similar and have much in common. For example, both perform high temperature, high pressure soxhlet-like extractions on up to 15 samples at a time. The kinetics of the extraction provide an extraction time of no more than...
This can occur with samples that are very low in fat. Moisture left in low-fat samples after extraction is the issue. As AOCS Procedure AM 5-04 indicates, when the Filter Bags are removed from the Extractor, they should be placed in an oven at 100...
For XT10, see Service Procedure Fault ETS (X0S002). For XT15, see Service Procedure Fault ETS (X5S016).
Possible Causes: (1) The dial on the heat sealer may not be turned up high enough. To seal the XT4 Bag, the dial must be set at approximately "6." (2) The heating element is broken. On the sides of the Teflon cover are two shiny metal strips that...
Below is a list of steps to verify that the XT instrument is operating correctly. Always refer to your manual for complete, detailed information. Refer to XT15 Service Procedure Solvent, Vent, and Water Diagram (X5S006) for a four-page diagram that...
When using the Mojonnier mixture of solvents, a black residue may build up on the bottom of the vessel. If this coating becomes too thick, an "Overtemp Fault" "<Enter>" may occur. The black residue can be cleaned using a drill with a long-shafted...
There are three primary reasons for high Crude Fat results. (1) The bag is not completely sealed; therefore the sample gets lost during extraction. (2) A bag picks up surface fat from another sample. (3) Incomplete drying in the pre-extraction phase...
The Overtemp message happens when the ANKOM HCl Instrument is heating but there is no solution in the vessel. The quartz heaters will heat the vessel and the ETS (Emergency Temperature Switches) will trip. Possible Causes: The instrument was not...
View the HCL Periodic Maintenance video .
To understand how to mix DE with various sample types, view the How to mix DE with samples video.
In developing our hydrolysis process, we did extensive research into acid and base hydrolysis. We found that the acid hydrolysis did well with a variety of samples, including those normally hydrolyzed in a base solution, i.e., dairy products. As a...