Instruments
Product Categories
Resources
FAQ Group
The MoistureStop Desiccant Pouch is a small, airtight zipper bag, utilized for desiccating all of the Filter Bags that are used with ANKOM instrumentation. A single pouch can hold a full run of F57, F58 or XT4 bags at one time. When folded, 6 IDF or...
Refer to the Fat Extraction Process section of the ANKOM XT10 Extractor Operator's Manual. See subsection, Sample Preparation Procedure for Crude Fat.
Yes. For instructions on how to extract liquid samples, view ANKOM XT10 & XT15 Crude (Free) Fat Extraction of Liquid Samples and refer to the Sample Preparation section of the Operator’s Manual.
Yes, all samples should be treated the same during the drying process. The biggest and most common cause of artificially high crude fat results is incomplete drying in the pre-extraction phase. Because our technique uses the indirect method of...
The XT instrument’s flowmeter should have a flow rate between 4-6 gallons per hour (If chilled coolant is used then the flow rate should be 2-3 gallons per hour). Note that this flow is intermittent and only turns on when cooling is required. When...
Crude Fat results could be low for any of the following reasons: The extraction was not run long enough. The sample was not post-extraction dried enough. Fat migrated from a bag during the pre-dry step. Desiccant was either not used or incorrectly...
Petroleum ether is the most common solvent recommended for use in XT Extractors. Many other solvents will work depending on your objectives but may require upgrades to the method or instrument for them to run effectively. Hexane, for example, will...
This is likely due to a valve that may be "stuck" closed. If the instrument has not been used in a while, it is possible that the solution is simply to clean the valves and clean any debris or buildup that may have settled on the valve piston. To...
There is a leak. The most common cause is a leaky Q1 valve, which would cause cooling water to flow into the vessel during the heating stage. To test: Turn the instrument completely off but keep the water supply on. Check to see if water is running...
The XT10 and XT15 look very similar and have much in common. For example, both perform high temperature, high pressure soxhlet-like extractions on up to 15 samples at a time. The kinetics of the extraction provide an extraction time of no more than...
This can occur with samples that are very low in fat. Moisture left in low-fat samples after extraction is the issue. As AOCS Procedure AM 5-04 indicates, when the Filter Bags are removed from the Extractor, they should be placed in an oven at 100...
For XT10, see Service Procedure Fault ETS (X0S002). For XT15, see Service Procedure Fault ETS (X5S016).
Possible Causes: (1) The dial on the heat sealer may not be turned up high enough. To seal the XT4 Bag, the dial must be set at approximately "6." (2) The heating element is broken. On the sides of the Teflon cover are two shiny metal strips that...
Below is a list of steps to verify that the XT instrument is operating correctly. Always refer to your manual for complete, detailed information. Refer to XT15 Service Procedure Solvent, Vent, and Water Diagram (X5S006) for a four-page diagram that...
When using the Mojonnier mixture of solvents, a black residue may build up on the bottom of the vessel. If this coating becomes too thick, an "Overtemp Fault" "<Enter>" may occur. The black residue can be cleaned using a drill with a long-shafted...
There are three primary reasons for high Crude Fat results. (1) The bag is not completely sealed; therefore the sample gets lost during extraction. (2) A bag picks up surface fat from another sample. (3) Incomplete drying in the pre-extraction phase...
High Fat Samples are defined as any meat sample with a fat content >20% and any plant sample with a fat content >40%. See XT Service Procedure High Fat Sample Handling (XTS014).
To perform an extraction of oilseeds, refer to Oilseed Extractions (FOSFA Equivalent).
Research with high fat samples has given us the following understandings: Some samples with fats that have low melting point triacylglycerides or emulsifiers may encounter fat migration during hydrolysis. To improve fat retention, increase the level...
Some samples with large amounts of carbohydrates (starches and cellulose) may bind small quantities of solvent and require additional post drying. To remove the bound solvent, dry at 100-105° C for one to three hours.
Check to ensure the procedures sent with your instrument were followed. Some possible variables that may affect your results are: General Problems: Neutral Detergent chemicals can come out of solution. Heat and stir to bring back into solution. The...
Rinse water can come directly from the tap. It does not have to be distilled.
The instrument is experiencing a pressure fault. There are three possible reasons for this fault. (1) More than 2 L of solution was in the vessel during the digestion process. A pressure fault will most commonly occur if there is more than 2 L of...
Refer to Service Procedure High Blank Bag Values (DES035).
The Blank Bag Correction allows for a weight adjustment based on actions that take place during solubilization. The resulting factor increases or decreases your sample bag weight. To determine the blank bag correction, include at least one empty...
When the neutral detergent solution gets cold, some of the components will begin to precipitate out of solution. As a result, the solution will turn milky white, and particles will be floating in it. These particles can be brought back into solution...
Depending on what you are testing, refer to section ADF Analysis, NDF Analysis, or Crude Fiber Analysis in the DELTA, A200, or A2000 Fiber Analyzer Operator's Manual.
For illustrations of melted trays, see A200 Service Procedure Temp Probe Replacement (FS016). Bag suspender trays can melt when there is not enough solution in the vessel to cover all of the trays while the instrument's heat is activated. Possible...
During a crude fiber or acid detergent procedure, the time frame during which samples are exposed to 100°C solution(s) is extremely important. If the Fiber Analyzer begins a procedure at a temperature higher than 20°C, the instrument will heat the...
It is possible to follow an NDF run with an ADF run using the same sample and Filter Bag. However, samples that are high in pectin, tannins, glucans, and fructans can produce lower-than-expected ADF values when NDF was run first. This is because...
F57 The porosity of the ANKOM F57 Filter Bag is 25 microns. The design of the filter bag provides a three-dimensional filter matrix similar to the filtering effect of a glass crucible. The F57 Filter Bag provides equal filtration without the manual...
Our grinding recommendation for samples to be extracted in our Fiber Analyzer is: 1 mm with a Wiley Mill 2 mm with a Cyclone Mill (such as UDY or Retsch)
To ensure that the controller settings are correct, refer to A200 Service Procedure Controller Reset Procedure (FS002) for instruments purchased pre-9/03, or, see A200 Service Procedure Controller Menu Setup (FS001) for instruments purchased post...
When the pressure in the vessel is too high, the pressure switch will turn off the heater. Possible Causes: There is too much solution in the vessel. Make sure two liters or less of solution is used in the vessel at a time. The motor is not turning...
Refer to Service Procedure Agitation System Check (DES031).
Test the rinse water with litmus paper during the last minute of the fourth hot water rinse. If the test indicates the presence of acid, repeat with fresh, hot water until a neutral pH is achieved.
Neutral detergent consists of varying amounts of 4 dry chemicals and triethylene glycol. If you remove only a portion of the dry chemicals to make less than 20L of solution, it is impossible to know if the solution has the correct ratio of chemicals...
Amylase is utilized to help digest the starch in the sample. Even using expired Alpha Amylase, it is highly likely the non-starch components, as well as most of the starch, are gone. Put the Filter Bags in a 1 L beaker with 80 °C water and 8 ml of...
With the lignin procedure, two common errors will create false high values. (1) The first error is placing the F57 Filter Bags into the 72% sulfuric acid when there is moisture in the bags and samples. To avoid this situation, after the ADF procedure...
The ANKOM fiber instruments can analyze for either Acid Detergent Fiber (ADF), Neutral Detergent Fiber (NDF), or Crude Fiber. NDF consists of hemicellulose, cellulose, and lignin. ADF consists of cellulose and lignin. Acid Detergent Lignin (ADL)...
It is important that the rinse water going into the instrument is at least 50°C for Crude Fiber and 70°C for detergent fibers. For safety purposes, most hot water heaters are set lower than 70°C. In addition, if there is a substantial distance...
Possible Causes: (1) The dial on the heat sealer may not be turned up high enough. To seal the XT4 Bag, the dial must be set at approximately "6." (2) The heating element is broken. On the sides of the Teflon cover are two shiny metal strips that...
Possible Causes: Motor wires are loose or disconnected. Unplug the instrument. Inspect the wiring to the motor. Contact ANKOM Technology if a loose or disconnected wire is found. If the wires are intact, run this test: Turn the instrument on and...
The F57 and F58 Filter Bags are ash-free and therefore can be used for ash determination. Refer to the Ashing Procedure Service Procedure (AS034) linked below.
Refer to Service Procedure Insert Samples Into F58 Bags (AS032).
To minimize bag material abrasion, use the new bag suspender (as of March 2015 - Part #: F11 ). It has been redesigned to include rounded bottom points on each tray.
Bag breakage can occur when the heat sealer setting is not hot enough to ensure a good seal. Make sure each bag is sealed within 4mm of its open end. This section includes a narrow polypropylene strip that provides elasticity to the heat-sealed area...
The instrument is experiencing a high-temperature fault. If your controller is set to 100°C, there are two possible reasons for this fault. (1) The agitation is not working properly during an assay. To check the agitation during an assay, remove the...