Skip to main content

Resource Center

Displaying 1 - 50 of 71
 
Frequently Asked Questions

We suggest a purity of 99.998% nitrogen or purer as it is used to expedite solvent evaporation. We have not tested less pure nitrogen and therefore cannot validate that it will not adversely affect analyte recovery.

 
 
Frequently Asked Questions

One ANKOM FLEX run includes four assays and will be completed in approximately two hours. Therefore, in an eight-to-ten-hour workday, the instrument can be run four times, resulting in sixteen assays per day, and eighty assays per five-day workweek...

 
Frequently Asked Questions

The FLEX can process up to 10g of sample for vitamin and cholesterol analysis. The sample size maximum is dependent on analyte content, analyte homogeneity within the sample, and sample composition. Samples that typically require 10g sample sizes are...

 
 
Frequently Asked Questions

No. The digestion vessels must not be left assembled for extended periods of time (e.g. overnight). If digestion vessels are left assembled for extended periods of time, the O-rings in the vessel bottom assemblies will deform and future integrity...

 
 
 
 
 
 
 
Frequently Asked Questions

Tips to avoid vessel leaks: (1) Do not spray water directly on the filter. Spray on the side of the vessel, so that it will flow onto the filter without force. You could also spray onto the mixer. (2) Always add the stir bar to the vessel before...

 
Frequently Asked Questions

Sorbent Disposal Bags ( Part #: FLEX-SDB ) hold spent sorbent from four SPE columns while allowing solvents to evaporate overnight in a hood. Dispose of the empty SPE columns and Sorbent Disposal Bags using your internal hazardous waste SOPs and SDS...

 
Frequently Asked Questions

The ANKOM FLEX Total Fat analysis run time is just under two and a half hours. Technician time is only five to ten minutes of set-up prior to running the instrument. This includes weighing samples into each digestion vessel and loading the digestion...

 
 
 
 
 
Frequently Asked Questions

This fault indicates a System Integrity Failure. One or more vessels did not reach or hold pressure properly. To resolve an MF1 fault: 1. Confirm the Low N2 setting is: 2.6 - 3.0 PSI. See Section 6.7.2 of the FLEX Operator’s Manual. 2. Tighten the...

 
Frequently Asked Questions

Maintenance is low. At the end of the day, the digestion vessel spray nozzles, digestion vessel bases, and SPE column top rings must be wiped clean with a dampened towel. Some O-rings require replacement generally after about two hundred runs. Refer...

 
 
 
 
Frequently Asked Questions

Possible Causes: (1) The level sensor has failed to detect the liquid that is in the vessel. If the vessel is full and this fault screen appears, press <STOP> to drain the fluid. You may want to capture the solution and re-use it. Refer to DELTA...

 
Frequently Asked Questions

The vessel temperature has exceeded the maximum limit. Possible Causes: (1) The solution level is low in the vessel, but the level sensor is indicating full. The level sensor has two main components that work together. These components are a Level...

 
 
 
 
 
 
 
 
 
 
 
 
Frequently Asked Questions

Possible Causes: The power cord is not plugged into the unit or power source. The main power switch is not turned on. One of the system circuit breakers (fuses) needs to be replaced. Refer to DELTA Service Procedure Fuse Replacement (DES025).

 
Frequently Asked Questions

Possible Causes: The rinse water supply temperature did not meet method standards. This may affect results with certain samples. Make sure the water supply has a temperature of no less than 50°C for Crude Fiber or no less than 70°C for ADF and NDF...

 
 
 
 
 
 
 
Frequently Asked Questions

In the event of a Drain Time-Out Fault (MF7) or an Abort during a Drain Module, users must flush the lines beneath the vessels to clear the residue prior to starting a new run. For more information, refer to Section 17 of the FLEX Operator’s Manual...

 
 
Frequently Asked Questions

Carry-over is extremely low (~0.08%) and therefore undetectable for most food and feed samples. When running concentrates, such as vitamin supplements, use a wash bottle to rinse the round bottom flask adapters with hexane prior to the next run to...