Frequently Asked Questions

1. Verify that you have a Gen2 Coordinator: A Gen2 Coordinator will have a number that is less than 1,000. If you have connectivity issues and your coordinator number is greater than 1,000 you have a Gen3 system. In that case, contact ANKOM using the...

 
 
Frequently Asked Questions

To update your GPM software: 1) Purchase the GPM software appropriate for your system. GEN2 systems (serial #’s <900) require GPM software version 11.4 ( Part #: RF3 ) GEN3 systems (serial #’s >1000) require GPM software version 15.0 ( Part #: RF3X )...

 
 
 
 
 
Frequently Asked Questions

The solenoid will become hot if the valve open check box is left checked. Normal valve openings are very brief and within a range of 10-1000 milliseconds, which does not cause the solenoid cylinder to get hot. If the Open Valves button is clicked or...

 
 
 
 
 
Frequently Asked Questions

Due to the discontinuation of certain parts, ANKOM can no longer replace the RF programmed chip (Part #: RF45) and the Circuit Board (Part #: 7007) in Gen2 modules. All other repairs and replacement parts will continue to be available indefinitely.

 
 
Frequently Asked Questions

Just prior to use, charge your rechargeable batteries (Part #: RF16 or RF46 ) for at least three hours using the 10-Station Battery Charger ( Part #: RF25 ). A fully charged battery when plugged into a module should read 6.7V or higher on the GPM...

 
Frequently Asked Questions

No, the batteries (Part #: RF16 or RF46 ) should not get hot. If this happens, the battery pack may be defective, or the module's circuit board may have failed. Install another fully charged battery pack in the module. Then, perform the module...

 
 
 
 
 
 
 
 
 
 
 
Frequently Asked Questions

Possible Causes: (1) The level sensor has failed to detect the liquid that is in the vessel. If the vessel is full and this fault screen appears, press <STOP> to drain the fluid. You may want to capture the solution and re-use it. Refer to DELTA...

 
Frequently Asked Questions

The vessel temperature has exceeded the maximum limit. Possible Causes: (1) The solution level is low in the vessel, but the level sensor is indicating full. The level sensor has two main components that work together. These components are a Level...

 
 
 
 
 
 
 
 
 
 
 
 
Frequently Asked Questions

Possible Causes: The power cord is not plugged into the unit or power source. The main power switch is not turned on. One of the system circuit breakers (fuses) needs to be replaced. Refer to DELTA Service Procedure Fuse Replacement (DES025).

 
Frequently Asked Questions

Possible Causes: The rinse water supply temperature did not meet method standards. This may affect results with certain samples. Make sure the water supply has a temperature of no less than 50°C for Crude Fiber or no less than 70°C for ADF and NDF...

 
 
 
 
 
Frequently Asked Questions

It is possible to follow an NDF run with an ADF run using the same sample and Filter Bag. However, samples that are high in pectin, tannins, glucans, and fructans can produce lower-than-expected ADF values when NDF was run first. This is because...

 
Frequently Asked Questions

"----" indicates there has been an electrical malfunction. You will need to reset the Controller. Refer to DELTA Service Procedure Controller Menu Setup (DES015). If the "----" still shows, you will need to replace the Ascon Controller ( Part #: A09...

 
Frequently Asked Questions

F57 The porosity of the ANKOM F57 Filter Bag is 25 microns. The design of the filter bag provides a three-dimensional filter matrix similar to the filtering effect of a glass crucible. The F57 Filter Bag provides equal filtration without the manual...