The Blank Bag Correction allows for a weight adjustment based on actions that take place during solubilization. The resulting factor increases or decreases your sample bag weight. To determine the blank bag correction, include at least one empty...
When the neutral detergent solution gets cold, some of the components will begin to precipitate out of solution. As a result, the solution will turn milky white, and particles will be floating in it. These particles can be brought back into solution...
Depending on what you are testing, refer to section ADF Analysis, NDF Analysis, or Crude Fiber Analysis in the DELTA, A200, or A2000 Fiber Analyzer Operator's Manual.
Yes, Electrostatic Discharge (ESD) can cause the flash programming of the XBee chips in your #RF1, RF1X , # RF5 , and RF5X modules to become corrupted and stop communicating with the RF GPM software. Therefore, we highly recommend using an ANKOM Anti...
To properly set up your instrument for the various fiber analyses, refer to the A2000 Operator's Manual.
The extraction process utilizes positive pressure at ~2.8psi.
We suggest a purity of 99.998% nitrogen or purer as it is used to expedite solvent evaporation. We have not tested less pure nitrogen and therefore cannot validate that it will not adversely affect analyte recovery.
No. The instrument exhaust lines (both waste and vent lines) are trapped and ready to be connected to an exhaust system.